DISPLAY MARK
CONDITION/CAUSE
CORRECTIVE ACTION
DESCRIPTION OF DISPLAY MARKS
Measurement is begun, but the pressure was in-
sufficient.
Change the pressurization setting one level by pressing “Pres-
sure Setting Switch” and measure once again. (Automatic
remeasuring will be done only 1 time.)
Appears when the blood pressure could not be
measured accurately. (Somenoise was detected or
moved arm during measuring.)
Repeat measurement again.
First exhaust air from the cuff, reapply the cuff properly and
remeasure.
Appears when the pulse could not accurately be
measured.
Replace the batteries.
Appears when the batteries become weak.
Appears when no memory is stored at pressing of
“Memory Recall Switch”.
Take a measurement.
Blood pressure will store automatically.
Units has a trouble.
Ask purchased shop.