background image

4.  Description of the instrument

1)   3 ½ digit LCD display

2)   OFF/ON button

3)   Measurement function selector

4)   Rubber case

5)   Hold button (save measurement value) 

6)   Output for testing cables

7)   Condenser socket

5.  Use

 

Warning 

1)  Always check before voltage measurement that the instrument is set to the 

correct measurement range and is not set to voltage or resistance or LED 

measuring. Make sure you always use the correct output for the kind of 

measuring that will take place.

2)  Be very careful when measuring voltage which exceeds 50 V. In particular, 

from measurement objects with a high output.

3)  Always check that the circuit is not powered before connecting testing 

cables, along with circuits (e.g. when measuring current).

4)  Make sure the circuit being tested is not powered before measuring 

resistance or buzzer/LED test. Make sure that the correct function and 

measurement range are selected. If you are unsure which is the correct 

measurement range begin with the highest and work your way down.

5)  You should be very careful when using the instruments on an inductive 

component. For example, transformers, relay coils or similar devices. A 

high voltage level may occur (be induced) at the measuring points as the 

electrical circuit is interrupted. 

6)  Make sure the testing cables are in good condition and that the insulation 

is not damaged.

7)  Make sure you do not exceed the overload limit indicated in the 

specifications. 

8)  Fuses to be changed must be of the correct type and correct value.

5.1 Measuring direct voltage and AC voltage

 

1)  Connect the black testing cable to the 

COM

 contact and the red testing 

cable to the 

VΩ

 contact.

2)  Set the function selector to 

  all measuring direct voltage or 

V ~

 all 

measuring AC voltage.

3)  Connect testing cables to voltage source or load to be measured.

5.2 Measuring direct current

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the

 mA 

contact (for measuring between 0–200 mA).

2)  Set the function selector to desired measurement Range 

 (Violet 

Range on measurement function selector).

3)  Connect in series testing cables between current sources and loading to be 

measured.

4)  When measuring current between 200 mA and 20 A (without fuse) follow the 

method above. But connect the red testing cable to the 20 A contact. When 

measuring high currents the instrument may only be connected a short while 

(15 seconds at 20 A). This is in order to avoid overheating the shunt in the 

instrument.

5.3 Measuring AC current

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the 

mA

 contact (for measuring between 0-200mA).

2)  Set the function selector to desired measurement range 

A~ 

(the little blue 

Range on measurement function selector).

3)  Connect in series testing cables between current sources and loading to 

be measured. When measuring current between 200 mA and 20 A (without 

fuse) follow the method above. But connect the red testing cable to the 

20 A

 contact. When measuring high currents the instrument may only be 

connected a short while (15 seconds at 20 A). This is done in order to avoid 

overheating the shunt in the instrument.

 

5.4 Resistance measurement

1)  Connect the black testing cable to the 

COM 

contact and the red testing cable 

to the 

VΩ

 contact.

2)  Set the measurement function selector to desired resistance range 

Ω

.

3)  Connect the testing cables to the circuits to be tested.

Warning: Check before measurement that the object to be tested is not 

powered. Maximum overload for the input: 250 V < 30 seconds under.

5.5 Diode test

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the 

VΩ

 contact.

2)   Set the measurement function selector on 

 on the blue range (the same 

as  ).

3)  Connect the black testing cable and red testing cable to anode (+) and 

cathode (-) on the diode to be tested.

4)  Record value for forward voltage loss on the display. If the display reads “1”: 

try reversing the polarity of the 

diode.

5.6  Measuring closed circuits  

 

(method to check breaks in a cable etc.)

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the 

VΩ

 contact.

2)  Set the measurement function selector on 

 (the same as  ).

3)  Connect the testing cables to the circuits to be tested. If the resistance is 

lower or almost the same as 60 Ω the buzzer is activated.

Warning: Check that the circuit to be tested is not powered. Maximum 

overload for the input: 250 V < 30 seconds.

5.7 Square wave generator

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the 

VΩ

 contact.

2)  Set the measurement function selector to desired frequency Range 

OUT

.

3)  Connect the testing cables to the component to be tested.

5.8 Capacitance measurement

1)  Set the measurement function selector to desired capacitance  .
2)  Insert condenser in condenser socket 

Cx

.

Warning: make sure the condenser being tested has been discharged.

5.9 Frequency measurement

1)  Connect the black testing cable to the 

COM

 contact and the red testing cable 

to the 

VΩ

 contact.

2)  Set the measurement function selector to desired frequency range 

Hz

.

ATTENTION:

 Never exceed limit values for voltage shown below.

Warning: Check that the circuit to be tested is not powered. Maximum 

overload for the input: 250 V rms < 30 seconds.

5.10 Save data (Hold)

The ”Save data” button (Hold) is used to store data during measurement. Press 

the button and you save the measurement value and 

H

 is displayed on the left-

hand side of the screen. The measurement value is erased when you press the 

button one more time.

6. Maintenance

Warning: To avoid electric shock, remove testing cables from powered 

measurement circuits before changing a battery or fuse.

6.1 Installing and changing battery/fuse

1)  Make sure the instrument is not connected to a measurement circuit. Turn 

the instrument off using the 

POWER

 button and remove testing cables from 

the outputs.

2)  Begin by removing protective casing.

3)  Unscrew the three screws to the back piece on the back of the instrument.

4)  Lift away back piece. 
5)  Change battery or fuse the type recommended under point 3. Specifications.

6)  Put on back piece and tighten the screws.

6.2 Cleaning

Wipe instrument regularly with a damp cloth coated in mild cleaning solution. Do 

not use abrasives or solvents. Make sure that no water gets into the instrument 

causing short circuits or other damage.

7.  Disposal

Follow local ordinances when disposing of this product. If you are unsure about 

how to dispose of this product contact your municipality.

GREAT BRITAIN

 For consumer contact, please visit www.clasohlson.co.uk and click on customer service.

1

2

3

7

4

5

6

Summary of Contents for BX3604W

Page 1: ...Omr de Uppl sning Noggrannhet Inimpedans verbelastningsskydd 200 mV 0 1 mV 0 8 1 10 M 250 V DC AC rms 2 V 0 001 V 0 8 1 600 V DC AC effektivv rde rms 20 V 0 01 V 200 V 0 1 V 600 V 1 V AC V Omr de Uppl...

Page 2: ...eln till V kontakten 2 St ll in m tfunktionsv ljaren p nskat resistansomr de 3 Koppla testkablarna till den krets som ska testas Varning Kontrollera innan m tning att m tobjektet som ska testas r str...

Page 3: ...V AC V Alue Erottelukyky Tarkkuus Taajuustoisto Sis ntuloimpedanssi 10 M 2 V 0 001 V 0 8 3 1 2 3 40 400 Hz Ylikuormitussuoja 600 V DC AC tehoarvo rms 20 V 0 01 V 200 V 0 1 V 600 V 1 V Mitta alueella...

Page 4: ...strumentin ylikuumeneminen 5 4 Resistanssimittaus 1 Liit musta testausjohdin COM kontaktiin ja punainen testausjohdin V kontaktiin 2 Aseta mittatoiminnonvalitsin halutulle resistanssialueelle 3 Liit t...

Page 5: ...ad protection 200 mV 0 1 mV 0 8 1 10 M 250 V DC AC rms 2 V 0 001 V 0 8 1 600 V DC AC rms value rms 20 V 0 01 V 200 V 0 1 V 600 V 1 V AC V Range Resolution Accuracy Frequency response Input impedance 1...

Page 6: ...g cable to the COM contact and the red testing cable to the V contact 2 Set the measurement function selector to desired resistance range 3 Connect the testing cables to the circuits to be tested Warn...

Page 7: ...0 M 250 V DC AC rms 2 V 0 001 V 0 8 1 600 V DC AC effektivverdi rms 20 V 0 01 V 200 V 0 1 V 600 V 1 V AC V Omr de Oppl sning N yaktighet frekvensgjengivelse Impedans inn 10 M 2 V 0 001 V 0 8 3 1 2 3 4...

Page 8: ...den r de til V V kontakten 1 Still inn m lefunksjonsvelgeren p nsket resistansomr de 2 Koble testledningene til den kretsen som skal testes Advarsel Kontroller at m leobjektet er str ml st f r m ling...

Reviews: