CleanSpace Respirators USER INSTRUCTIONS
22
English
English
6.
Leave the contaminated area immediately if any of the following conditions occur:
Any part of the system is damaged
Airflow into the mask decreases or stops
The battery or filter alarms are triggered (even if only the audible or only the visible alarm triggers)
Breathing becomes difficult
You feel dizzy or your vision is impaired
You smell or taste contaminants
Your face, eyes, nose or mouth become irritated
You suspect the concentration of contaminants may have reached levels at which this respirator may no longer
provide adequate protection.
7.
You should have specific exiting and decontamination procedures as part of your workplace respiratory protection
program. Follow those procedures, together with the instructions below when removing the respirator.
To remove the respirator, find the Mask Release Button
(see Section 5) and press to release the mask from the
blower.
The mask release button has a rough, dimpled
finish that you can feel with your fingertips. All the other
buttons are smooth.
Note: The respirator motor will automatically switch off
and go into Standby Mode in approximately 10 seconds
when no breath is detected’
DO NOT REMOVE THE RESPIRATOR UNTIL YOU HAVE
VACATED THE CONTAMINATED AREA unless you have pressing health reasons to do so (for instance you are experiencing
dizziness and believe removing the respirator while you leave the contaminated area may help)
11.
Working in Noisy Environments
In most noisy environments the CleanSpace Respirator alarms are still clearly audible to the wearer. But if there is a chance
that you will not hear the alarms, you must take the following extra precautions
Never enter the contaminated zone unless all three green battery indicator lights are illuminated
Never work in the high noise area for more than four (4) hours. At the end of four hours, leave the contaminated zone
and check that
-
The filter blocked light is not illuminated; AND
-
All three green battery indicator lights are still illuminated
If two or fewer green battery indicator lights are illuminated, recharge CleanSpace Respirator until 3 lights are showing
again
If the filter blocked light is illuminated, change the filter
Be particularly aware of difficulty breathing or of the air flow stopping. If these things happen, exit the contaminated
zone immediately.
12.
Changing the Filter
Before using CleanSpace Respirator you must ensure that you are using the correct filter type for the environment you are
working in. The CleanSpace Respirator blower is supplied with a particulate filter.
IT IS ESSENTIAL THAT THE CORRECT FILTER TYPE IS SELECTED FOR THE CHOSEN APPLICATION.
Summary of Contents for PAF-0060
Page 468: ...www cleanspacetechnology com...