Cleaver Scientific Thermal Cycler
7
protocol run. Keeping the heated lid open during a protocol run might cause
“HEATER ERROR”, because the heated lid fails to reach the set temperature.
3.
To evaporate the condensation water inside sample block, and to avoid “HEATER
ERROR”, users should turn off the Lid Temperature under SYSTEM mode before
heating the sample block and keeping the lid open.
3.4
Lid Opening/Closing
There are two ways to open the lid: 1) Slide the lid to the rear, 2) lift up the lid. The first
method is recommended for common operations and to prevent the skin burn from the
heated lid. The second method is used to clean up the heating plate of the lid.
Rotate Lid Knob in a clockwise direction to lower down the heating plate and press down
on the cap of tubes, tighten until you feel resistance. This adjustment will provide better
contact between the tube or plates and the block for better performance. Rotate the Lid
Knob in counter-clockwise direction as shown in Figure 4 until the Lid pops out. Only
when the Lid pops out, users can slide the Lid to the rear as shown in Figure 5.
To lift up the lid for heating plate cleaning, please refer Figure 6 ~ 8.
Figure 4. Rotate Lid Knob to open the lid