background image

 

 

Page 16 

 

Section 5 Troubleshooting Guide

 

 

Problem 

Cause 

Solution 

Bands  sharp  but  not  enough  bands 
seen 

Gel agarose percentage too high 
Incomplete digestion 

Decrease agarose percentage. 
Review  enzyme  activity,  digest 
further. 

Band smearing and streaking 

Agarose has improper endosmosis 
 
Salt concentration in sample too high 
Excessive power and heating 
 
Sample spilled out of well 
 
 
Incomplete 

digestion, 

nuclease 

contamination, bad enzyme 
Sample  wells  cast  through  the  gel. 
Sample  leaks  along  bottom  of 
running surface 

Consult  Cleaver  Scientific  about 
agarose. 
Reduce salt concentration to ≤0.1M. 
Reduce  voltage.  See  electrophoresis 
instructions. 
Apply  sample  carefully.  Increase  gel 
thickness  for  large  sample  volumes. 
Adjust comb height. 
Heat 

sample. 

Review 

enzyme 

activity. Digest sample further. 
Comb should be placed to 1 to 2 mm 
above  the  base  of  the  running 
surface. 

Curved line or distortion of bands 

Bubbles in sample wells 

Remove 

bubbles 

prior 

to 

electrophoresis. 

Curved bands, smiles 

Sample overload 

Reduce load. 

Differential relative mobilities 

Sample spilled out of wells 
 
Unit not leveled 

Samples should have proper density. 
Apply carefully. 
Level unit. Use a steady work bench. 

Gels crack 

Too high  voltage gradient, especially 
with  low  melting  temperature 
agarose or low gel strength gels 

Reduce  voltage.  Run  gel  at  lower 
temperature. 

High  MW  bands  sharp;  low  MW 
bands smeared 

Gel agarose percentage too low 

Increase  agarose  percentage.  Switch 
to polyacrylamide. 

Ragged bands 

Sample density incorrect 
Sample well deformed 
 
 
 
 
Excessive power or heating 

See sample application instructions. 
Carefully  remove  comb,  especially 
from  soft  gels.  Make  sure  gel  has 
solidified. 
Cooling  soft  gels  aids  in  comb 
removal. 
Reduce  voltage.  See  electrophoresis 
instructions. 

Slanted lanes (bands) 

Gel not fully solidified 
Comb warped or at an angle 

Gel to solidify for at least 30-45min. 
Check alignment of comb. 

 
 
 
 
 

Summary of Contents for multiSUB MSCHOICE

Page 1: ...MSMINI MSMIDI MSCHOICE MSCHOICEST MSMAXI Instruction Manual Catalogue No MSMINI7 MSMINI10 MSMINIDUO MSMIDI7 MSMIDI10 MSMIDIDUO MSCHOICE7 MSCHOICE10 MSCHOICE15 MSCHOICETRIO MSCHOICEST20 MSCHOICEST25 MS...

Page 2: ...2 Setting up the Horizontal Gel Tanks 9 3 3 Gel Preparation 10 3 4 Gel Pouring 11 3 5 Running the Gel 13 3 6 Gel Staining and Viewing 14 3 7 Solutions 14 Section 4 References 15 Section 5 Trouble Sho...

Page 3: ...uld read the complete manual thoroughly The unit must never be used without the safety lid correctly in position The unit should not be used if there is any sign of damage to the external tank or lid...

Page 4: ...L CAB MSMINIDUO MS7 UV7 MS7 UV10 MS7 UVDAM Pack of 2 2 x MS7 8 1 1mm 8 sample MS7 LG Strips MS7 WP Platform CSL CAB MSMIDI7 MSMIDI10 MSMIDIDUO Units include tank lid and electrodes and include the fol...

Page 5: ...x L MS15 UVDAM Pack of 2 2 x MS15 20 1 1mm 20 sample MS15 LG Strips MS15 WP Platform CSL CAB MSCHOICETRIO MS15 UV7 MS15 UV10 MS15 UV15 MS15 UVDAM Pack of 2 2 x MS15 20 1 1mm 20 sample MS15 LG Strips M...

Page 6: ...x L MS20 UVDAM Pack of 2 2 x MS20 20 1 1mm 20 sample MS20 LG Strips MS20 WP Platform CSL CAB MSMAXIDUO MS20 UV10 MS20 UV20 MS20 UVDAM Pack of 2 2 x MS20 20 1 1mm 20 sample MS20 LG Strips MS20 WP Platf...

Page 7: ...s include dishwashing liquid Hexane and Aliphatic hydrocarbons The units should not be left to in detergents for more than 30 minutes The units should never come into contact with the following cleani...

Page 8: ...ltitude 2 000m Temperature range between 4 C and 65 C Maximum relative humidity 80 for temperatures up to 31OC decreasing linearly to 50 relative humidity at 40OC Not for outdoor Use This apparatus is...

Page 9: ...o that there is no gap between the lid and the leading edge of the cable fitting 4 Refit the lid Instructions for fitting Loading Guides These can be fitted to enhance visibility of the wells if desir...

Page 10: ...multiSUB Maxi Tray 20 x 10 cm 20 x 15 cm 20 x 20 cm Gel volume for a 5mm thick gel 100 mL 150 mL 200 mL 2 Add the agarose powder to a conical flask 3 Add the appropriate amount of 1x TAE or TBE soluti...

Page 11: ...gel leakage 2 Place the comb s in the grooves Each tray has more than one comb grove so that multiple combs can be used Using multiple combs increases sample number available per gel but decreases ru...

Page 12: ...to the edge of the gel and dispersed using a pipette tip 5 Allow the agarose to set ensuring that the gel remains undisturbed 6 Carefully remove the gel casting gates and comb and transfer the gel inc...

Page 13: ...ally 3ul of sample buffer is adequate but less may be used with sample volumes of less than 10ul 2 Fill the unit with buffer until the gel is just flooded with buffer This will give the fastest resolu...

Page 14: ...d water again ensuring the gel is completely immersed 3 Rinse the gel twice for a couple of seconds with distilled water 4 Transfer the gel to a UV Transilluminator 5 The samples will often appear as...

Page 15: ...of 50 glycerol 0 25 bromophenol blue and 0 25 xylene cyanole FF in 1x TAE buffer Only 1 10 ml of the 10x loading dye should be prepared Ethidium Bromide Solution Add 10 mg of Ethidium Bromide to 1 ml...

Page 16: ...the running surface Curved line or distortion of bands Bubbles in sample wells Remove bubbles prior to electrophoresis Curved bands smiles Sample overload Reduce load Differential relative mobilities...

Page 17: ...ng Platform MSMINICP Cool pack and Platform MSMINIBSB Buffer Saver Blocks pk 2 saves 100ml of buffer MS7 UVS 7cm UV Gel Scoop CSL CAB Electrophoresis cable Black Red MSMIDI7 multiSUB Midi 10 x 7cm UV...

Page 18: ...ms MS15 PE Positive Electrode MS15 NE Negative Electrode MSMS15 LG Adhesive Loading Guides MS15 WP Viewing Platform MSCHOICECP Cool Pack and Platform MS15 20 FC multiSUB Choice Maxi Flexi caster MSMAX...

Page 19: ...7 16 0 75 16 MS10 16 0 75 16 MS15 16 0 75 25 MS20 25 0 75 20 MS10 20MC 0 75 18 MS15 18MC 0 75 30 MS20 30 0 75 20 MS15 20 0 75 36 MS20 36 0 75 28 MS15 28MC 0 75 40 MS20 40MC 0 75 30 MS15 30 0 75 multiS...

Page 20: ...5 16 1 5 25 MS20 25 1 5 20 MS10 20MC 1 5 18 MS15 18MC 1 5 30 MS20 30 1 5 20 MS15 20 1 5 36 MS20 36 1 5 28 MS15 28MC 1 5 40 MS20 40MC 1 5 30 MS15 30 1 5 multiSUB Mini 2mm multiSUB Midi 2mm multiSUB Cho...

Page 21: ...supplies as 20 blister packs of 10 x 0 5g tablets CSL RUNSAFE CSL RUNSAFE Package 1 ml vial CSL TBEP Powdered Tris Borate EDTA Running Buffer 10 sachets 1litre pk TBE10X1L Cleaver Buffer Tris Borate...

Page 22: ...dified Units which have accessories or repaired parts not supplied by CSL or its associated distributors have invalidated warranty CSL cannot repair or replace free of charge units where improper solu...

Reviews: