background image

 

MEDIPULSE 2342se ERGONOMIC SUCTION BLAST CABINET 

Page 25

 

 

 

 

 

 

© 2019 

CLEMCO INDUSTRIES CORP.  

  www.clemcoindustries.com  

  Manual No. 24939, Rev. C, 03/19

 

8.10 

Blow-back through media hose  

 

8.10.1

  Blockage  in  nozzle.  Remove  the  nozzle  and 

check for blockage. 
 

8.10.2 

Air jet may be too large for nozzle. Refer to the 

table in Figure 2. 
 
 

8.11 

Blasting does not begin when the foot pedal 

is pressed. 

 

8.11.1 

Door 

interlocks 

not 

engaging. 

Check 

adjustment, per Section 5.5. 
 

8.11.2 

Blocked  or  leaking  control  lines.  Check  all 

urethane tubing for blockage or leaks. 

 
8.11.3 

Foot  pedal  valve  malfunction.  Check  foot  pedal 

alignment, and inlet and outlet lines for pressure. 
 

8.11.4 

Make  sure  lines  are  not  reversed  on  the  foot 

pedal or pilot regulator. Refer to the cabinet controls and 
plumbing schematic on Page 33, Figure 44. 

 
8.11.5 

Pressure  regulator  may  be  set  to  low  or  OFF. 

Adjust blast pressure, per Section 5.1. 
 

8.11.6 

Make  sure  the  air  compressor  is  operating  and 

air-supply valves are open.

 

 

8.11.7

  Check the nozzle to see if it is plugged. Refer to 

Section 8.4.

 

 
 
8.12 

Blasting continues after releasing foot pedal

 

 

8.12.1

  Make  sure  the  3-way  valve  in  the  foot  pedal 

exhausts  air  when  the  pedal  is  released.  If  it  does  not 
exhaust,  check  the  inbound  air  line  for  blockage,  if  no 
blockage, replace the valve. 

 
 
8.13 

Media buildup in cabinet hopper; media does 

not convey to reclaimer

 

 

NOTE:  Do  not  pour  media  directly  into  the  cabinet 
hopper, as overfilling may occur.

 Overfilling will result 

in  media  carry  over  to  the  dust  collector  and  possible 
blockage in the conveying hose.

 

 
8.13.1

  Exhauster  motor  rotating  backward.  The  motor 

should rotate as indicated by the arrow on the exhauster 
housing.  If  it  does  not  rotate  in  the  proper  direction, 

lockout

 and 

tagout

 electrical power and switch the motor 

leads, as shown on the motor plate. Refer to the system’s 
wiring schematic. Refer to Sections 2.4. 
 

8.13.2

  Dust-collector  damper  closed  too far, restricting 

air  movement  through  cabinet.  Adjust  static  pressure, 
per Section 5.4. 

 
8.13.3

  Dust-collector  filter  cartridge  clogged.  Refer  to 

Section  5.6  to  adjust  pulse  pressure;  refer  to  Section 
6.2.4 for additional cartridge pulsing. 

 
8.13.4

  Hole worn in flex hose between cabinet hopper 

and  reclaimer  inlet  or  between  the  reclaimer  outlet  and 
dust-collector  inlet.  Inspect  hoses  and  replace  them  as 
needed. 
 

8.13.5

  Reclaimer  door  open.  DO  NOT  operate  unless 

door is closed. 
 

8.13.6

  Obstruction  in  flex  hose.  Remove  hoses  and 

check for blockage. 

 

 

8.14 

Static shocks 

 
8.14.1 

Cabinet and/or operator not grounded. Abrasive 

blasting generates static electricity. The cabinet must be 
earth-grounded  to  prevent  static  buildup.  See  Section 
2.3.  If  shocks  persist,  the  operator  may  be  building  up 
static. Attach a small ground wire, such as a wrist strap, 
from the operator to the cabinet. 

 
8.14.2

  Gloves wearing thin. Inspect gloves and replace 

them as needed. 

 
8.14.3 

Avoid  holding  parts  and  blasting  off  the  grate. 

Static  will  build  in  the  part  if  not  dissipated  through  the 
metal  cabinet.  If  blasting  parts  off  the  grate  cannot  be 
avoided, attach a ground wire between the cabinet and 
the part. 
 
 

8.15 

Dust leaking from cabinet 

 

8.15.1

  Refer to Section 8.13. 

 
 

8.16 

Dust leaking from dust collector

 

 
8.16.1

  Cartridge  not  seasoned;  season  cartridge,  per 

Section 7.10.

 

 

8.16.2 

Damaged  or  loose  cartridge.  Inspect  filter 

cartridge and tighten or replace as needed.  
 

8.16.3 

Faulty  seal  on  the  dust  collector  cone.  Inspect 

seal and replace if damaged. 
 
 

Summary of Contents for MediPulse 2342se

Page 1: ...ible that the products described in this material may be combined with other products by the user for purposes determined solely by the user No representations are intended or made as to the suitabili...

Page 2: ...ates a hazardous situation that if not avoided could result in minor or moderate injury WARNING Warning indicates a hazardous situation that if not avoided could result in death or serious injury DANG...

Page 3: ...PLACEMENT PARTS 9 0 Optional Accessories field installed 9 1 BNP Gun and Feed Assembly 9 2 Cabinet Assembly 9 3 LED Light Assembly 9 4 View Window 9 5 Metering Valve 9 6 Foot Pedal 9 7 Filter Regulato...

Page 4: ...6 1 Unless otherwise specified at the time of purchase the cabinet is provided with a No 5 5 16 orifice ceramic nozzle and No 5 5 32 orifice air jet Optional more durable tungsten carbide and boron c...

Page 5: ...formance from the reclaimer When using larger nozzles the maximum mesh size of media will be smaller than normally recommended Using media finer than those recommended may decrease visibility and at s...

Page 6: ...g 1 2 HP 115 230V 1 PH 60 HZ motors wired for 115 volts FLA 20 7 208 3 4 240 3 5 1 10 2 A power cord with u ground plug is provided 2 0 INSTALLATION 2 1 General Installation Notes 2 1 1 Select a locat...

Page 7: ...ssories 2 3 Ground the Cabinet 2 3 1 To prevent static electricity buildup attach an external grounded wire from an earth ground to the grounding lug on the left rear of the power module 2 4 Connect E...

Page 8: ...lacement 3 2 Cabinet Curtains Installation 3 2 1 Match curtains to corresponding wall and doors 3 2 2 Front and rear walls Position the curtain on the wall to be protected Using the curtains as templa...

Page 9: ...flow particle weight and size Reclaimer static pressure is set by adjusting the outlet damper refer to Section 5 4 to adjust static pressure Refer to Section 5 8 for manometer instructions 3 5 Reclai...

Page 10: ...hown by removing the fitting s compression nut sliding it over the end of the tubing inserting the tubing into the fitting and tightening the nut onto the fitting 3 6 6 Refer to the differential press...

Page 11: ...pen the fill door and stir media until it starts to flow Before the container becomes too heavy to easily manage replace the plug and empty the container Replace the plug when the reclaimer is empty o...

Page 12: ...immediately WARNING Shut down the cabinet immediately if dust discharges from the dust collector or cabinet Check to make sure the dust collector filter cartridge is correctly seated and that it is no...

Page 13: ...adjust pressure unlock the knob by pulling it out as shown in Figure 14 and turn it clockwise to increase pressure or counterclockwise to decrease pressure Pressure may drop slightly from closed line...

Page 14: ...h finer media higher with coarser media Run the media through several blast cycles allowing the reclaimer to function with these settings Inspect media in the reclaimer and fines in the dust collector...

Page 15: ...imum pulse pressure of 90 psi is attained and additional pulsing as described in Section 6 2 4 does not increase visibility or decrease differential pressure replace the cartridge per Section 7 9 5 7...

Page 16: ...will cause carry over of good media to the dust collector 5 8 6 2 To take occasional readings Leave the needle protector on the needle and insert the needle into the unused end of the tubing The ends...

Page 17: ...require more maintenance to the cabinet s media recovery system Friability of media Media that rapidly breaks down require more maintenance to the cabinet s media recovery system and dust collector 6...

Page 18: ...e and after each use Figure 23 6 2 4 Additional cartridge pulsing The cartridge is pulsed each time the foot pedal is pressed or released Additional pulsing should be performed per the following instr...

Page 19: ...n the inside of the cabinet To replace loosen the clamps with a screwdriver replace the gloves and tighten the clamps 7 1 3 Quick change type clampless installation Quick change gloves are held in pla...

Page 20: ...media leakage around the view window or if gaskets are worn or otherwise damaged Figure 25 7 5 2 Remove the window and window frame per Section 7 6 7 5 3 Remove all the old gasket material and clean t...

Page 21: ...e gasket to bond Repeat the process for each side compressing the ends to seal 7 7 3 Diffuser lens replacement 7 7 3 1 Remove the four nuts and washers that attach the light mount cover to the cabinet...

Page 22: ...plate against the inner wall of the reclaimer 7 8 9 New self drilling sheet metal screws are provided with the wear plate while forcing the wear plate against the reclaimer wall install the screws thr...

Page 23: ...e container and drape it over the top edge Reattach the container to the lid and latch the lock ring making sure the lid and clamp are secure Replacement liners are shown in Sections 9 11 7 9 12 Seaso...

Page 24: ...1 9 Obstruction in flex hose between the cabinet hopper and reclaimer inlet Inspect hose for blockage 8 1 10 Paddle wheel worn Check wheel for wear 8 2 Abnormally high media consumption 8 2 1 Door on...

Page 25: ...mer mounting bolt may help to prevent bridging of fine mesh media NOTE To avoid the possibility of compressing media a vibrator should be set up to start when the foot pedal is pressed and stop when t...

Page 26: ...as indicated by the arrow on the exhauster housing If it does not rotate in the proper direction lockout and tagout electrical power and switch the motor leads as shown on the motor plate Refer to the...

Page 27: ...k alignment If the switch does not engage the pulse when the cam follower rides onto the lobe the switch assembly is too far from the lobe Loosen the two screws on the bottom of the pedal slide the sw...

Page 28: ...spect the foot pedal per Sections 8 17 2 8 17 3 and 8 17 4 Figure 32 8 18 A steady stream of air is heard within the dust collector when the foot pedal is not pressed 8 18 1 Cam follower does not ride...

Page 29: ...tion 9 2 Retaining ring wide spray 12038 Nozzle nut wide spray 11916 or Nozzle guard wide spray 12295 Time delay door locks for both doors 24163 Tumble basket 2 gallon 12227 Automatic pulse kit 21773...

Page 30: ...1914 Urethane covered 11574 2 O ring 08975 3 Gun body 11802 4 Lock nut air jet 11913 5 Rubber sleeve 12097 6 Air jet assembly with Item 5 No 4 12342 No 5 12343 No 6 12344 7 Fitting hose 3 8 NPT x 1 2...

Page 31: ...66 11 Clamp for 4 hose 11577 12 Fitting 1 8 NPT elbow x 1 8 barb 11733 13 Regulator 1 8 NPT pilot 12715 14 Gauge pressure 1 8 NPT cbm 01908 15 Foot pedal with tubing 20194 16 Grommet media air hose 3...

Page 32: ...ck No 1 Window glass 12 5 x 19 5 laminated 12212 2 Gasket 5 16 x 3 4 applied to cabinet per foot 6 feet required 00189 3 Gasket 5 32 x 3 4 applied to window frame per foot 6 feet required 00192 4 Cove...

Page 33: ...w x 1 8 barb 11733 17 Tie nylon wire 12139 Figure 42 9 8 Filter Regulator Inlet Piping Assembly Figure 43 Item Description Stock No Filter regulator inlet piping assembly Includes all items shown 2152...

Page 34: ...1 2 flare 11351 19 Hose end 1 2 barb x 1 2 female swivel 15002 20 Hose end 1 2 barb x 3 8 male NPT 06369 21 Blow off nozzle 13116 22 Air hose 1 2 specify ft required 12472 23 Gun assembly BNP No 5 st...

Page 35: ...Stock No Repair kit diaphragm pulse valve 21600 1 Valve 1 diaphragm pulse 19578 2 Valve 3 way 12202 3 Fitting 1 8 NPT straight x 1 8 barb 11732 4 Nipple 1 8 NPT hex 01962 5 Actuator air pilot 19123 6...

Reviews: