PULSAR
®
PLUS 55-S SUCTION BLAST CABINET
Page 19
© 2021 CLEMCO INDUSTRIES CORP.
www.clemcoindustries.com
Manual No. 30423
2. Press the left "off" (arrow down) button until "OF"
appears briefly on the LED display.
3. To decrease OFF time (for example, to change pulse
OFF time from 40 seconds to 20 seconds), press the
left "off" (arrow down) button until 20 seconds shows
on the display.
4. To increase OFF time (lengthen time between pulses
from 20 seconds to 40 seconds), press the right
"on/arrow up" until 40 seconds shows on the display.
When the new OFF time appears on the display do not
press anything; after a few seconds, the display starts
flashing, indicating the new time is being saved. Once
the new time is saved, the unit resumes operation with
the new time setting.
TEST
of
f
on
on
OFF TIME
Figure 23
NOTE: Do not adjust ON time. Adjusting ON time will
consume additional air but will not improve cartridge
cleaning. If ON time is accidentally changed, reset it to
factory settings, per Section 5.9.3.
5.9.3 Reset pulse ON time to factory setting, 0.15
seconds (150 milliseconds).
NOTE: There is no need to change ON time. ON time is
factory set for 150 milliseconds (0.15 second), as shown
in Figure 22. If ON time is accidentally changed, reset it
to 150 milliseconds as follows:
NOTE: The numbers shown in Figure 24 signify the
following step numbers:
1. Make sure the ON subrange is set to milliseconds, as
noted in Section 5.9.5.2, Figure 26.
2. Press the right "on/arrow up" button and ON appears
briefly on the display.
3. If the number on the display is above 15, decrease
time by pressing the left "off/arrow down" button until
15 appears.
4. If the number on the display is less than 15, increase
time by pressing the right "on/arrow up" button until
15 appears.
When 15 appears on the display do not press anything;
after a few seconds, the display starts flashing,
indicating the new time is being saved. Once the new
time is saved, the unit resumes operation with the new
time.
on
TEST
of
f
on
ON TIME FACTORY SETTING
Figure 24
5.9.4 Reset timer function to factory "A"
setting
NOTE: There is no need to change the function setting.
If the function is accidentally changed, reset it to the
factory setting as follows:
1. Turn OFF the power switch located on the cabinet
control console.
2. Press and hold down the timer "TEST" button and
Switch ON the power switch.
3. Release the "TEST" button after the current timer
function (A through F) appears on the display.
4. Use the "off/arrow down" or "on/arrow up" buttons to
select function "A".
hrs
min
sec
x10ms
1
2
4
3
LED
Display
hrs
min
sec
x10ms
1
3
2
4
LED
Display