27
-
When entering the phone number, entering (
) represent a 3-sec. pause. (e.g. Switchboard
system, extension, etc.)
-
Up to 20 digits including (
) are allowed for each telephone number.
-
The 1st, 2nd, 3rd, and 4th Tel. numbers will be used as primary medical alarm number,
secondary medical alarm number, primary fire/CO2 alarm number, and secondary fire/CO2
alarm number, respectively.
11# -14# Account Number Programming (01-04):
-
Set the account numbers for the 1
st
-
4
th
Tel. numbers used for Reporting
respectively
.
-
In each field, key in an
Account Number
of 4 or 6 or 8 digits.
After editing and updating, click write to apply new settings.
4.4.1.4 Settings
The “Settings” tab allows you to program general configuration for CTC-1052-QT-DT35AW.
Click “Write” when you have completed all settings to confirm.
31# Ringer Enable for Incoming Calls:
-
When disabled, CTC-1052-QT-DT35AW will operate silent for an incoming call.
-
When enabled, CTC-1052-QT-DT35AW will sound a ring tone as an audio alert for an incoming
call. The user can answer the incoming call by pressing the HELP BUTTON or Pendant Button.
32# Audible Fault Alert:
-
When turned on, CTC-1052-QT-DT35AW will sound warning beeps every 30 seconds for AC
Failure.
-
When turned off, CTC-1052-QT-DT35AW will NOT sound warning beeps for a fault situation.
34# Emergency Button Enable:
-
When this function is programmed as
Button On
, the HELP BUTTON is activated. When
pressed, it will send an Emergency alarm report to the Monitoring Station.
-
When this function is programmed as
Button Off
, the HELP BUTTON is deactivated.
35# Silent Mode:
-
When Silent Mode is On, all normal operation warning beeps, guard time tone, key tone and
rings will be silenced. Speech Reporting made under Silent Mode will automatically be switched