Step 3
Immerse cable in detergent (second time)
A
Fill a sink or container with enough fresh pH 7–11 enzymatic detergent to
completely immerse the cable.
B
Place the cable in the sink or
container so that all parts and
surfaces of the cable are immersed
in the detergent.
COVER
WITH
C
Remove cable from the sink or container and drain the sink or container of
pH 7–11 enzymatic detergent.
Continue to Step 4 on page 42
Step 2
Brush cable under cold running water (first time)
Using a soft-bristled brush, brush the
entire length of the cable under cold
running water.
Pay particular attention to
connectors, crevices, and difficult-to-
clean areas.
UNDER
41
Main cleaning: Full manual cleaning