Step 2
Inspect cable
A
Use a magnifying glass to inspect the cable for any remaining soil. Pay
particular attention to
crevices, grooves and difficult-to-clean areas.
Always repeat the entire cleaning process, including initial manual
cleaning, if any soil is present during the inspection stage.
INSPECT
FOR SOIL
Continue to part B of Step 2 on page 58
57
Preparation for sterilisation