Preparation for sterilisation
Step 2
Inspect cable (continued)
B
1. Check the connector is securely attached to the cable.
2. Use a magnifying glass to inspect the cable for damage. Check for:
●
Faulty or cracked insulation
●
Broken electrical wires
●
Crushed or kinked cable
●
Damaged connector
If a Versius Electrosurgery Instrument Cable shows signs of defects or damage,
contact CMR Surgical Customer Services (see page 7).
CRACKED
INSULATION
BROKEN WIRE
KINKED CABLE
CRACKED
INSULATION
DAMAGED
CONNECTOR
DAMAGED
CONNECTOR
Continue to Step 3 on page 59
INSPECT
FOR DAMAGE
58