Equipment required
Steam steriliser
with dynamic
air removal (prevacuum) cycle, in
compliance with ISO 17665-1 and
with the following parameters:
Sterilisation tray
Prevacuum steam sterilisation parameters
Parameter
Value
Temperature
134–137 °C (273–279 °F)
Exposure time
3–18 minutes (minimum holding time)
Dry time
20 minutes (minimum)
OR
Parameter
Value
Temperature
132 °C (270 °F)
Exposure time
4 minutes (minimum holding time)
Dry time
20 minutes (minimum)
65
Sterilisation