Step 3
Leave cables to cool down
1. Remove the sterilisation tray from the
steriliser after the cycle finishes.
2. Leave the cable to cool down before continuing to Step 4.
Allow electrosurgery instrument cables to cool down gradually to room
temperature before handling. Sudden changes in temperature could
damage electrosurgery instrument cables.
Step 4
Transport and store cables
Transport and store the required cables in individual steam-permeable
sterilisation peel pouches to point of use in line with local guidelines or
hospital protocol.
Transport and store cables in line with local guidelines.
Take care when handling and transporting Versius Electrosurgery
Instrument Cables to avoid damage.
Reprocessing is complete
67
Sterilisation