Procedures Overview
Main cleaning: Full manual cleaning
Step 1
Immerse cable in detergent (first time)
Step 2
Brush cable under cold running water (first time)
Step 3
Immerse cable in detergent (second time)
Step 4
Brush cable under cold running water (second time)
Step 5
Rinse cable under running critical water
Main cleaning: Automated cleaning and thermal disinfection
Step 1
Place cable into washer/disinfector tray
Step 2
Load tray into washer/disinfector
Step 3
Run washer/disinfector
Step 4
Remove cable from washer/disinfector
18
Reprocessing Versius Electrosurgery Instrument Cables