Step 1
Moisten cable
do ONE of the following options
Choose either
Option A
or
Option B
to keep the cable moist.
Alternative methods may be used, according to local guidelines.
Begin point-of-use preparation within 30 minutes of the surgical
procedure ending.
Option A:
using liquid solution
Place the cable into the
container, ensuring that it is lying flat and fully covered
by liquid. Choose from either:
premixed pH 7–11
enzymatic detergent
(preferred solution)
cold water
OR
COVER IN
OR
Continue to Step 2 on page 26
24
Point-of-use preparation