Point-of-use preparation
Step 2
Transport cable
1. Seal the container with the cable
inside.
2. Transport the cable with care,
in line with local guidelines or
hospital protocol.
Start initial manual cleaning within
4 hours of finishing point-of-use
preparation.
SEAL IN
Continue to Initial manual cleaning on page 29
TRANSPORT
26
Point-of-use preparation