Initial manual cleaning
Step 1
Remove cable from container
u
Start initial manual cleaning within 4 hours of finishing point-of-use
preparation.
1. Remove the cable from the container.
2. Drain the container of the enzymatic detergent or water.
Step 2
Brush cable
Using a soft-bristled brush, brush the
entire length of the cable under cold
running water to remove excess soil.
Pay particular attention to
connectors, crevices, and difficult-to-
clean areas.
Continue to Step 3 on page 33