Initial manual cleaning
Step 4
Rinse cable under cold running water
Rinse the cable under cold running water for a minimum rinsing time of
1 minute 30 seconds.
RINSE UNDER
1 MINUTE
30 SECONDS
MINIMUM
Continue to page 35
34
Initial manual cleaning
Summary of Contents for versius
Page 3: ......
Page 4: ...4...
Page 6: ...6...
Page 14: ...14...
Page 15: ...15 Reprocessing Versius Electrosurgery Instrument Cables...
Page 20: ...20...
Page 27: ...27...
Page 28: ...28...
Page 36: ...36...
Page 43: ...43...
Page 44: ...44...
Page 51: ...51...
Page 52: ...52...
Page 61: ...61...
Page 62: ...62...
Page 71: ...71...