Main cleaning: Full manual cleaning
Step 1
Immerse cable in detergent (first time)
A
Fill a sink or container with enough fresh pH 7–11 enzymatic detergent to
completely immerse the cable.
B
Place the cable in the sink or
container so that all parts and
surfaces of the cable are immersed
in the detergent.
COVER
WITH
C
Remove the cable from the sink or container and drain the sink or container
of the detergent.
Continue to Step 2 on page 41
40
Main cleaning: Full manual cleaning