Main cleaning: Full manual cleaning
Step 4
Brush cable under cold running water (second time)
Using a soft-bristled brush,
brush the entire length of
cable under cold running
water.
Pay particular attention to
connectors, crevices, and
difficult-to-clean areas.
UNDER
Step 5
Rinse cable under running critical water
Rinse the cable under running critical water for a minimum rinsing time of
1 minute 30 seconds.
RINSE UNDER
1 MINUTE
30 SECONDS
MINIMUM
Continue to Preparation for sterilisation on page 53
42
Main cleaning: Full manual cleaning