Sterilisation
Step 1
Load sterilisation tray into steam steriliser
Load the sterilisation tray into the
steam steriliser according to the steam
steriliser manufacturer’s instructions.
LOAD
INTO
Step 2
Run steriliser
Set parameters and run sterilisation cycle (according to ISO 17665-1):
Prevacuum steam sterilisation parameters
Parameter
Value
Temperature
134–137 °C (273–279 °F)
Exposure time
3–18 minutes (minimum holding time)
Dry time
20 minutes (minimum)
OR
Parameter
Value
Temperature
132 °C (270 °F)
Exposure time
4 minutes (minimum holding time)
Dry time
20 minutes (minimum)
SET AND RUN
Continue to Step 3 on page 67
The sterilisation process should
conform with ISO 17665-1.
66
Sterilisation