7
Introduction
Introduction
This manual provides reprocessing instructions for Versius Electrosurgery Instrument
Cables including cleaning, disinfection and sterilisation, and equipment recommended
by CMR Surgical Ltd.
These instructions must be read and fully understood before reprocessing Versius
Electrosurgery Instrument Cables.
These instructions have been validated by CMR Surgical. This does not preclude the
statutory requirements of the sterile services department to undertake process
monitoring, testing and quality control to ensure the effectiveness of the cleaning,
disinfection and sterilisation procedures.
Versius Electrosurgery Instrument Cables are supplied non-sterile, and must be
reprocessed before first use.
Discolouration of metal surfaces is possible when alkaline enzymatic detergents are
used (pH 8–11). However, this does not impair the functional capability of the Versius
Electrosurgery Instrument Cables.
If there are any questions regarding these instructions or if any Versius Electrosurgery
Instrument Cables show signs of defects or damage, please contact CMR Surgical
customer services. To arrange the return of any affected Versius Electrosurgery
Instrument Cables, please contact CMR Surgical technical support.
CMR Surgical Limited,
1 Evolution Business Park,
Milton Road, Cambridge CB24 9NG, UK
Tel: +44 (0) 1223 750 975
For technical support:
If your system requires technical support
or service, please call our technical support
line:
+44 (0) 1223 750 975
Email:
customer.service@cmrsurgical.com
CMR Surgical Limited,
1 Evolution Business Park,
Milton Road,
Cambridge CB24 9NG, UK
www.cmrsurgical.com
Summary of Contents for versius
Page 3: ......
Page 4: ...4...
Page 6: ...6...
Page 14: ...14...
Page 15: ...15 Reprocessing Versius Electrosurgery Instrument Cables...
Page 20: ...20...
Page 27: ...27...
Page 28: ...28...
Page 36: ...36...
Page 43: ...43...
Page 44: ...44...
Page 51: ...51...
Page 52: ...52...
Page 61: ...61...
Page 62: ...62...
Page 71: ...71...