Sampling Data Logger User’s Manual

 

CO2Meter.com

 

 

 

PAGE 

13

of 

17

 

1%

 

or

 

30%

 

Sampling

 

Data

 

Logger

 

Calibration can be performed using either 100% CO2 calibration gas (typically nitrogen, argon, etc.), or using a 
fresh source of air, assumed to be approximately 400ppm. 

NOTE

: Refer to the GasLab

®

 User’s Manual for complete instructions.  

 
Calibration Procedure  

To calibrate your meter, follow these steps: 

1.

 

Turn ON the meter. 

2.

 

Expose the meter to ambient air (assumed to be at 400ppm) or supply calibration gas bottle/cylinder 
(100% nitrogen or argon) with the appropriate demand regulator. 

3.

 

Wait 25 seconds to collect a sample. Write down this value as the “before” value. 

4.

 

Click the 

CALIBRATE

 button after selecting your calibration gas. 

5.

 

Wait 25 seconds again. This time, the meter will take a sample and use this data to adjust zero values. The 
displayed measurement will show the new calibration value (0 or 400ppm, depending on method used). 

6.

 

Disconnect the Calibration gas and wait 25 seconds. All displayed data will now be based on the new 
calibration value. Repeat this procedure if the sensor is still operating outside of its specified range or if 
readings vary greatly. 

5%,

 

60%,

 

and

 

100%

 

Sampling

 

Data

 

Logger

 

These models do not feature Automatic Background Calibration (ABC) algorithm due to the CO2 scale they 
measure. To ensure the highest accuracy, we recommend calibrating these units with calibration gas (Nitrogen for 
ZERO or CO2), close to the concentration being measured. Alternatively, a 100% or ambient calibration can be 
performed.  

To perform a calibration, attach the unit to your computer, power it with the wall adapter and either expose it to 
atmosphere or supply it with your calibration gas, fed with a demand‐based regulator. 

Open the GasLab

® 

software on your personal computer and click the 

CONFIGURE

 

SENSOR

 button. Click the 

TURN

 

PUMP

 

ON

 

CONTINUOUSLY

 button to ensure continuous flow. (To apply any changes, the meter has to be 

restarted). Make sure to turn this back off after calibration. 

Zero

 

or

 

Fresh

 

Air

 

Calibration

 

Supply gas and select the appropriate concentration in the screen. Click the 

CALIBRATE

 button. The sensor reading 

should instantly reflect the calibration. 

High

Concentration

 

Calibration

 

Write down the original Zero Value before adjustment for future reference. Apply the desired concentration 
calibration gas; adjust the Zero Value in increments of 10 pressing the large 

SAVE

 button on the right until the unit 

displays the correct concentration. 

 

 

 

Summary of Contents for CM-0001

Page 1: ...M 0002 CM 0003 CM 0050 CM 0056 See Table 1 on page 7 for details WARNING Install our complimentary GasLab software before connecting your CO2Meter product s to your computer Failure to do so may affec...

Page 2: ...mplimentary GasLab software before connecting your CO2Meter product s to your computer Failure to do so may affect the ability for GasLab to detect your meter If this happens please follow the instruc...

Page 3: ...INSTRUMENT OVERVIEW 8 LCD DISPLAY 8 QUICK START GUIDE 9 POWERING THE METER 9 OPERATION 9 CONNECTING THE METER TO YOUR PC 9 DATA STORAGE 10 DATA LOGGING 10 SETTINGS 11 DISPLAY SETTINGS 11 PORT SETTINGS...

Page 4: ...community CO2Meter Inc appreciates your business and looks forward to working with you and your team in the future Please take some time to read through this manual in order to become familiar with th...

Page 5: ...Optional Accessories If the meter is to operate in environments where humidity is very high 95 RH an Extreme Moisture Filter CM 0103 can be purchased separately This filter will allow free flow of sa...

Page 6: ...T is not supported on Media Center or Tablet editions Installer will optionally install NET Framework Visit www co2meter com pages downloads to download our complimentary GasLab software to your compu...

Page 7: ...d on user selected time intervals Scientific devices such as this sampling meter require users to have intimate knowledge of the meter s features limitations operation required software and specificat...

Page 8: ...percentage format for all other models Logging ON OFF Percentage of free memory Display Shows CO2 concentration level data logging and free memory space Power Jack Connect included power supply Power...

Page 9: ...iod of time batteries should be removed to prevent damage due to potential battery leakage The POWER switch on the left side of the meter turns the unit ON OFF The LOG switch on the right side of the...

Page 10: ...fer to the GasLab User s Manual for more information Data Logging These models feature an internal memory capable of storing data when not connected to a computer This allows you to simultaneously rea...

Page 11: ...ere you can change the date and time on the meter for more accurate reference on the records The default time of your meter could vary depending on your location Review these settings before you start...

Page 12: ...jolted or otherwise mechanically disturbed the sensor can drift requiring recalibration All calibration procedures follow a single point calibration routine that effectively shifts the zero point of...

Page 13: ...or if readings vary greatly 5 60 and 100 Sampling Data Logger These models do not feature Automatic Background Calibration ABC algorithm due to the CO2 scale they measure To ensure the highest accurac...

Page 14: ...other models General Warm up Time 1 min instant measurements Life Expectancy 15 years Maintenance Interval No maintenance required Data Logging Data Points 60 000 Programmable Interval Data Date time...

Page 15: ...set For more troubleshooting tips on GasLab software see its manual located at www co2meter com pages downloads USB Driver Installation Instructions To install the appropriate USB port drivers compati...

Page 16: ...jected to misuse neglect accident modifications or repairs by you or by a third party No employee or reseller of CO2Meter com s products may alter this warranty verbally or in writing Liability All li...

Page 17: ...7 of 17 Contact Us We are here to help For information or technical support please contact us support co2meter com 386 256 4910 Technical Support 386 872 7665 Sales www co2meter com Address CO2Meter I...

Reviews: