HANDBOOK (V
4.0)
COMPLETE CONTROL SYSTEM GEN2
31
CALIBRATION UNDO
can be performed using the
Complete Calibrate
button on the
prosthesis, too. Simply hold it down for 6 seconds (until you hear a double beep and see
the LED flashing blue) and let it go. This will jump the calibration back to what it was just
before the most recent calibration activity.
CALIBRATION RESET
can be performed using the
Complete Calibrate
button on the
prosthesis, as well. To reset ALL calibrated motions using the Coapt button, hold it pressed
for about 10 seconds (until you hear it play a short melody) then release. This will reset the
system back to a starting state and turn the button’s LED magenta. Don’t expect control
of the prosthesis when the LED is magenta—the control memory is empty and it will have
to get at least one calibration to operate a prosthesis again.
Calibrating control on-screen (via mobile or desktop)
In case you would prefer to calibrate your Coapt system by watching on-screen prompts, great calibration tools
are provided in the Coapt
Complete ControlRoom
software interface or app.
To use the PC software, make sure that you have an available Coapt
Complete Communicator
Bluetooth radio
dongle, and then make sure you download and install the Coapt
Complete ControlRoom
application software
from the Coapt website. To use the iPhone/iPad application, visit the Apple store for the Coapt
Complete
ControlRoom
app.