HANDBOOK (V
4.0)
COMPLETE CONTROL SYSTEM GEN2
45
INPUTS PANEL ELEMENTS
EMG INPUT BARS
Dynamic signal activity on eight horizontal bars of color represent the eight EMG
channel inputs of the
Complete Control
System Gen2. The on-screen colors
correspond to the EMG wire colors of the EMG Interface Cable.
A real-time signal quality assistant monitors for EMG electrodes losing contact with
user’s skin (“lift-off”). When lift-off is detected, a warning message is displayed on the
corresponding signal channel.
EMG INPUTS -
ON/OFF TOGGLES
Each EMG Input channel has an on/off toggle that will determine its inclusion or
exclusion from the set of pattern recognition inputs. It is strongly suggested that all
eight EMG inputs are used to ensure optimal performance.
SCROLLING
OUTPUTS DISPLAY
Below the set of EMG inputs is a scrolling display that shows real-time control outputs
from the current state of the
Complete Control
system’s calibrated pattern
recognition classifier. No motions will be displayed when the system is in a Reset state.
For convenience, visibility of this output display can be toggled on or off by clicking
the “eye” icon.
SNAPSHOT -
BUTTON
(CAMERA ICON)
The camera icon opens a pop-up window that contains a snapshot of the EMG Inputs
activity at the moment the button was pushed. The pop-up allows for the captured
imaged to be saved for later use.