HANDBOOK (V
4.0)
COMPLETE CONTROL SYSTEM GEN2
47
UNDO –
BUTTON
(COUNTER-
CLOCKWISE
ARROW ICON)
Reverts the system to the calibrated state it was in just before the most recent
calibration activity, that activity either being single-motion or full-sequence
calibration.
The Undo functionality only applies to one calibration step; that is, undoing more than
one calibration step is not possible.
RESTORE POINTS –
BUTTON
(BOOKMARK
ICON)
Opens a pop-up window for storing or loading a Calibration Restore Point.
SETTINGS –
MENU
(COGWHEEL ICON)
Opens a pop-up window for
Complete Control
calibration settings:
CALIBRATION PACE
-
SELECTOR
Sets the pace at which prompts will be spaced during a
calibration sequence.
ADAPTIVE
ADVANCE
TM
-
TOGGLE
When “On” (default), enables the
Adaptive Advance
TM
continuous-learning algorithms to meld any new calibration
data into the existing state of users’ control (see
Adaptive
Advance
TM
section for more).
CONTROL COACH
®
LED SENSITIVITY -
SELECTOR
Adjusts when the
Control Coach
®
white LED on the
Complete
Calibrate
button will come on. For example, when set to 2 stars,
the
Complete Calibrate
button indicator will only come on
when the overall
Control Coach
®
star rating is a 2 or less (see
Control Coach
®
section for more).
SINGLE-CLASS
CALIBRATION DATA
AMOUNT -
SLIDER
Determines how much calibration data users will be tasked to
provide in order to “fill-the-bucket” during any Single Class
Calibrations.
RESET ALL -
BUTTON
Sets the
Complete Control
system to a blank starting state. The
control memory will be emptied and will need at least one new
calibration to actuate the prosthesis.