5)
Type
T<enter>
to access the Test Menu.
6) Type
S5,5<enter>
to view the output of the Pressure Sensor. Y ou may initially see -
1,000,000.00, but you should see regular readings within a few seconds as in the following:
> s5,5
Note: Hit any key to halt output.
BP
-1000000.0000
1009.6403
1009.6403
1009.6403
1009.6403
1009.6403
1009.6403
7) Compare the readings from the pressure sensor with the readings from your reference
barometer. T here should be no m ore than 0.05 m b difference between the ZENO
®
3200
barometer and the laboratory reference barometer. If the ZENO
®
3200 barometer is determined
to be out of tolerance, it must be sent to the manufacturer for recalibration or it must be replaced.
If you have any other questions or require assistance with your ZENO
®
3200 system, please
contact the Customer Service Department: