What you need:

  •  Accutrend Plus
  •  Test strips & related code strip
  •  Lancing device e.g. Safe-T-Pro  
    Plus
  •  Alcohol wipe
  •  Gloves
  •  Tissue

1.

Place the instrument on a level, 
vibration-free surface and switch on 
by pressing the On/Off button.

2.

Check the display
If no code has been stored in the in-
strument the display will look like this

Otherwise the last stored code will be 
displayed & the screen will look like 
this:

 

The meter will store the data of one 
code strip per test parameter. 
To check other tests & code numbers 
in the meter press the set key. 

3.

If your test/code is not displayed it 
will be necessary to insert the code 
strip for the test strip lot you wish to 
use.
Hold the code strip behind the black 
bar and with the chamber flap closed 
insert the code strip as far as it will 
go and remove it immediately.  Do 
this firmly with consistent speed.
If successful you will hear 1 beep.
If unsuccessful the screen will look 
like this

Lift the chamber flap to clear the 
message or wait a few seconds and 
try again.

4.

Remove a test strip from the 
container and close the container 
immediately.  Check it’s appearance 
for discolouration.  Do not used if 
discoloured. 
Insert the test strip firmly with 
consistent speed as far as it will go.  
If successful you will hear 2 beeps.  
A flashing arrow indicates that the 
meter is now ready for you to open 
the chamber flap to apply the drop of 
blood.  At this stage keep it closed.  
You have approx 2 minutes before 
the meter will auto-off.

5.

Choose the lancing site (pref middle 
finger, outer side) and clean with 
alcohol wipe.  Dry with tissue.

6.

Lance the finger and milk site to 
produce a drop of blood.  

Wipe off 

the first drop.

  Milk the site again to 

form a large hanging drop of blood.

7.

Lift the chamber flap and remove 
the test strip (leave the flap open).  
Apply the hanging blood directly to 
the reaction pad.  Do not touch the 
pad.  The application area must be 
completely covered with blood.

8.

Insert the test strip back into the 
instrument.

                                                                                                                                                      
                                                                                                                                                      
                                                                                                                                                      
                                                                                                                                                      
    

9.

Close the chamber flap.  This starts 
the measurement & countdown.

Test 

Time

Glucose

12s

Cholesterol

180s

Triglycerides

174s

Lactate

60s

The meter will beep 4 secs out from 
displaying the result.

10.

Interpretation of Hi or Lo results

Test

Lo

Hi

Glucose

<1.1 mmol/L

>33.3mmol/L

Cholesterol

<3.88 mmol/L >7.76mmol/L

Triglycerides <0.8mmol/L

>6.86mmol/L

Lactate 
(blood)

<0.8mmol/L

>21.7mmol/L

11.

When measuring tests on several 
people clean and disinfect the 
instrument with an alcohol wipe 
before each measurement.

12.

Accutrend

®

 Plus - Quick Reference Guide

Note:

A plausibility check must be performed for all

 glucose

 measurements.  See page 80 of the user’s manual for instructions.

   
   

Protection against infections: 

 There is a potential risk of infection.  Medical staff and other persons using the Accutrend  

    Plus instrument to perform measurements for more than one patient must be aware that any object coming into contact    
    with human blood is a potential source of infection.