2
NUCLEUS® 7 HYBRID MODE PROFESSIONALS GUIDE
Introduction
Nucleus Hybrid mode is suitable for most recipients who have aidable post-operative
residual hearing.
In Hybrid mode, the Nucleus 7 Sound Processor
delivers synchronised signals:
• acoustic signals via the ear canal
• electric signals to the implant.
Hybrid mode is enabled by:
1. Attaching a Hybrid receiver to a Nucleus 7
Sound Processor, and
2. Programming it in the Custom Sound
software.
Your role as a clinician
1.
Check Cochlear implant compatibility:
Hybrid mode, as part of the Nucleus system,
is compatible with a range of Cochlear implants. Contact Cochlear for the latest
information on compatible implants.
2.
Check candidacy:
Perform audiometric testing and counselling to determine if the
recipient is a candidate for Nucleus Hybrid mode (see
Candidacy
on page 4).
3.
Fit the Hybrid system to the recipient:
Select the correct Hybrid receiver and dome
to fit the recipient’s ears (see
Fit the Hybrid system
on page 6).
4.
Assemble the Hybrid system:
Install the Hybrid receiver and dome and show the
recipient how to wear the system (see
Assemble the Hybrid system
on page 8).
5.
Program the Hybrid system:
Add a Hybrid MAP using the Custom Sound software
(see
Program Hybrid mode
on page 14).
6.
Counsel the recipient
on using and caring for their Hybrid system (see
Care and
maintenance
on page 16).