Potential for Radio and Television Interference
SLS PCS generates and uses radio frequency energy, and if not installed and used
properly, that is, in strict accordance with the
manufacturer’s instructions, may cause
interference to radio and television reception. It has been type tested and found to
comply with Class A emission limits for a computing device in accordance with the
specifications in Subpart J of Part 15 of FCC Rules, which are designed to provide
reasonable protection against such interference when operating in a commercial
environment. SLS PCS is not intended for use in a residential Class A environment.
SLS PCS requires a medical power/ground. If your SLS does cause interference to radio or
television reception, you are encouraged to try to correct the interference by one or more of
the following measures:
Reorient the receiving antenna
Relocate SLS PCS with respect to the receiver
If necessary, you should consult Codonics Technical Support or an experienced
radio/television technician for additional suggestions. You may find the following booklet
prepared by the Federal Communications Commission helpful:
How to Identify and Resolve
Radio-TV Interference Problems
. This booklet is available from the U.S. Government Printing
Office, Washington, D.C. 20402, Stock No. 004-000-00345-4.
This product is in conformity with the protection requirements of EC Council directive
MDR 2017/745/EU(CE) on the approximation of the laws of the Member States relating to
medical devices. This product satisfies the Class A limits of IEC 60601-1-2 for Professional
Healthcare Facilities and CISPR 11. A declaration of conformity with the requirements of the
Directive has been signed by a Codonics vice president.
Guidance Regarding Electromagnetic Emissions and Immunity
Suitable environments are as follows:
SLS550i is intended for use in hospital and clinical environments including operating
rooms and the perioperative environment.
SLS550i has not been evaluated for use near HF surgical equipment. If use near HF
surgical equipment is desired, the user is responsible for verifying proper operation
of the SLS550i. If SLS550i does not perform correctly in this environment, move the
SLS550i farther from the source of the electromagnetic disturbance.
SLS550i has not been evaluated for use in emergency medical vehicles or in residential
applications.
Summary of Contents for SLS550i
Page 8: ......
Page 31: ...4 Insert the SmartDrive 5 Insert the output bin Startup ...
Page 37: ...1 At the Login prompt scan your user badge barcode or manually enter your user ID ...
Page 55: ......
Page 62: ...Elektromagnētisks Imunitāte Standarti un Pārbaude Līmeņi ...
Page 79: ...4 Ievietojiet SmartDrive 5 Ievietojiet izlaide atkritumu tvertne Uzsākt ...