15. If the optical disc position is too far left or right, adjust 

the 

Bottom Drive Tray 

calibration value to move the 

robotic arm left and right. Reselect the 

Pick

 or 

Move 

button to verify the new location.

16. Press the 

right arrow

 button or 

left arrow

 button until 

Top 

Drive Tray 

appears as the 

Component.

17. Press the 

Drop

 button to have the robotic arm place the 

optical disc in the top optical drive tray.

18. Verify that the robotic arm is placing the optical disc as 

shown in step 4. If calibration is needed, perform steps 6 
through 14.

fåéìí=_áå

The following steps define how to calibrate the picking 
position from the input bins.

1. Press the 

right arrow

 button or 

left arrow

 button until 

Left 

Bin 

appears as the 

Component

.

2. Press the 

Move

 button to have the robotic arm hold the 

optical disc at the top of the bin.

3. If the optical disc position is too far left or right, adjust 

the value to move the robotic arm right and left. Reselect 
the 

Move 

button to verify the new location.

4. Press the 

Drop

 button to drop the optical disc into the left 

bin.

5. Press the 

Pick

 button and carefully watch the optical disc 

position relative to the bottom of the bin. If the optical 
disc is not centered, average the optimal top and bottom 
position for the final bin calibration setting. 

6. Repeat steps 1 through 5 on the right bin.

mêáåíÉê=qê~ó

The following steps define how to calibrate the picking 
position from the printer tray.

1. Press the 

right arrow

 button or 

left arrow

 button until 

Printer Tray 

appears as the 

Component

.

2. Press the 

Drop

 button to have the robotic arm place the 

optical disc in the printer tray.

3. If the optical disc position is too far left or right, adjust 

the calibration value to move the robotic arm right and 
left. Reselect the 

Pick

 or 

Move 

button to verify the new 

location.

4. If the optical disc position is too far forward in the printer 

tray and coming close to the two disc stops (shown 
below), this could indicate that the printer tray is being 
pulled back by the disc tab retention spring. If the printer 
tray is pulling back, the optical discs could become 
dislodged when the printer tray is retracted. Please contact 
Codonics Technical Support for further instruction.

NOTE: 

If unable to achieve center, bias the optical disc 

towards the right of center. 

NOTE: 

As a final check, it is recommended to move the 

robotic arm to another location and return back to the 
bottom optical drive. Complete a final check of optical 
disc location by selecting the 

Drop

 button on the 

Bottom 

Drive Tray 

component. Having the robotic arm move 

from and back to the calibration location ensures the 
robotic arm has accurately read the encoder position.

NOTE: 

Press 

Pick

 button to remove the optical disc from 

the bottom optical drive before proceeding to the next 
step. 

NOTE: 

As a final check, it is recommended to move the 

robotic arm to the right bin and then back to the left bin. 
Complete a final check of optical disc location by 
selecting the 

Move

 button corresponding to 

Left Bin

NOTE: 

Pressing the 

Move

 button will lower the robotic 

arm and hold the optical disc in the printer tray. 

NOTE: 

If unable to achieve center, bias the optical disc 

towards the right of center. 

NOTE: 

As a final check, it is recommended to move the 

robotic arm to the right bin and then back to the printer 
tray. Complete a final check of optical disc location by 
selecting the 

Drop

 button corresponding to 

Printer Tray

.

NOTE: 

Another way to check for printer tray pull back 

is to manually pull the printer tray all the way out and 
let go. No pull back should be observed.

Summary of Contents for Virtua Robotic Arm Utility

Page 1: ...the top cover and place a folded piece of paper or card stock in the optical sensor as shown in the following image This will allow access to the drives and print tray while the cover is open 3 Go to...

Page 2: ...the Recorder Pressing the left arrow button will decrease the value and move the robotic arm to the left as it would appear from the front of the Recorder 10 The Robotic Arm Cycle Test see below can...

Page 3: ...steps 1 through 5 7 If both optical drives need the same adjustment mark the location of the drive bay bracket on the chassis using a felt tip pen at the front of the Recorder as shown 8 Remove all op...

Page 4: ...calibration value to move the robotic arm right and left Reselect the Pick or Move button to verify the new location 4 If the optical disc position is too far forward in the printer tray and coming c...

Page 5: ...g the pick depth pick force 5 Use the Drop and Pick button to view how forcefully the robotic arm presses into that component adjust the force using Step 4 until satisfied 6 Repeat step 1 through 5 on...

Page 6: ...the pick height for the print tray and DVD drives This process may be required if the pick height calibration on the GUI was attempted without a disc in the print tray or DVD drive tray 1 Connect a US...

Page 7: ...minutes to complete 1 To begin the test press the Start button in the Robotic Arm Cycle Test section A Robot Test Running message and flashing green light will display on the dashboard A yellow light...

Reviews: