Setting Temperature Range
The Tempinspect USB data logger has 7 temperature range presets.
These define the upper and lower temperature limits for the device's
temperature range alarm system.
To set the temperature range, press the multifunction as follows:
2 quick presses followed quickly by a 5 second press to set 2
°
c
to 8
°
c range (Default)
3 quick presses followed quickly by a 5 second press to set -1
°
c
to -10
°
c range
4 quick presses followed quickly by a 5 second press to set -2
°
c
to -8
°
c range
5 quick presses followed quickly by a 5 second press to set 2
°
c
to 25
°
c range
6 quick presses followed quickly by a 5 second press to set 15
°
c
to 25
°
c range
7 quick presses followed quickly by a 5 second press to set 15
°
c
to 30
°
c range
8 quick presses followed quickly by a 5 second press to set -15
°
c
to -30
°
C range
After selecting the range, immediately press and hold the multifunction
button for 5 seconds to confirm your selection. The LED light will blink
red as many times as required by your selection to confirm your
settings. (Example: 4 red blinks confirm -2
°
c to -8
°
C range)
•
Start/Stop
STEP1
--,
pressfor5
seconds
0
_
--.,.....J
STEP2
--,
0
press and
ho1dfor
Ssaconds
...,J
STEP3
--,
,
J
operation (temperature
o
logging) quickly press
for 5 seconds and
+
quickly re-press for
another 5 seconds and release the multifunction button.
Green
and
Red
leds will flash alternatingly during 5 seconds, indicating that the
Temptrail has been initialized. Once started, the device operation can be
terminated at any time by repeating the same procedure to start the
operation.
The operation cannot be reset. The data logger has a 10 minute start
delay.
•
Check status
At any time, a quick press on multifunction button informs
the tempinspect
´
s state:
Before initiating operation, the
tempinspect is IDLE, which is
represented by a single flash of the
green
led, followed by a single flash of
the
red
led.
During operation, the tempinspect is
RECORDING, which is represented by
four flashes of the
green
led.
After the operation is terminated, the
tempinspect is DONE, which is
represented by two flashes of the
red
led, followed by two flashes of the
green
led.
•
Check alarm
STEP 1
Short press the
..,._
Multifunctional
Button
Tempinspect will show the visual alarms
via a red or green flashing led light
•
by
by pressing the status button once. If
the
products have been exposed over the
chosen temperature range, a red led
light will flash one time for every interval equivalent of exposure time
over the chosen temperature range, and if products are exposed under
the chosen temperature range, a red led light will double flash for every
interval temperature reading equivalent. Therefore if you set the data
logger to take temperature with 1 0 minute intervals, the led alarms will
give one flash for every ten minutes over the temperature range or
double red led Flash for every ten minutes under the Temperature Range.
If no out of range exposure the green led will flash four times confirming
data logger is recording and no alarms have taken place. The below is an
example for a set measurement time every ten minutes. The visual
alarms may be seen prior to stopping the data logger by a simple press
of the status button. once the data logger is stopped, connect it to your
usb drive and a pdf and csv file are automatically generated with all
required information.
STEP2
i
STEP3
J
Flash Once
Flash Twi ce
Quickly
For every 10 minutes ABOVE the
maximum temperature limit, the
red
led
will flash once.
For every 10 minutes BELOW the
minimum temperature limit, the
red
led
will flash twice quickly
•
Detailed Diagnosis
Get the detailed diagnosis of the temperature at which the cargo has
been exposed.
fr
"'
"
0
J
Remove
The Top
The Delay may be customi
z
ed via our custom software available
for download on www.coldchainvision.com
/
software.
To start the Tempinspect
led in dic ator
For resetting multi-use data logger press status button
9
times and on tenth
time keep pressed for 5 seconds. Green and red light will do simultaneously
a long flash indicating data logger is resetting.
Once reset the data logger will give 1 green light and 1 red flash indicating
that now the logger is ready to be programmed as new.