1-2 Polystat
Section 1
Cole Parmer
Many refrigerants which may be undetectable by human senses are heavier than air and will
replace the oxygen in an enclosed area causing loss of consciousness. Refer to the chiller's
nameplate and the manufacturer's most current MSDS for additional information.
Never connect the process fluid inlet or outlet fittings to your building water supply or
any water pressure source.
Never operate the chiller without fluid in the reservoir.
Other than water, before using any fluid, or when performing maintenance where contact
with the fluid is likely, refer to the manufacturer’s MSDS and EC Safety Data sheet for
handling precautions.
Ensure, that no toxic gases can be generated by the fluid. Flammable gases can build up
over the fluid during usage.
Do not use automotive antifreeze. Commercial antifreeze contains silicates, or any
Organic Acid Technologies (OATs), that can damage the pump seals. Use of automotive
antifreeze will void the manufacturer’s warranty.
Never use corrosive or flammable fluids with this chiller. Use of these fluids will also void
the manufacturer’s warranty.
When using a process fluid mixture of ethylene glycol and water or propylene glycol and
water, check the fluid concentration and pH on a regular basis. Changes in concentration
and pH can impact system performance. See Section 3.
Never operate damaged or leaking equipment.
Never operate the chiller or add fluid to the reservoir with panels removed.
Do not clean the chiller with solvents, only use a soft cloth and water.
Drain the chiller before it is transported and/or stored in, near or below freezing
temperatures. Store the chiller in the temperature range -25°C to 60°C (with packaging),
and <80% relative humidity.
Always turn the chiller off and disconnect the supply voltage from its power source
before moving the chiller or before performing any service or maintenance procedures.
Transport the chiller with care. Sudden jolts or drops can damage its components.
Refer service and repairs to a qualified technician.
Performance of installation, operation, or maintenance procedures other than
those described in this manual may result in a hazardous situation and will void the
manufacturer's warranty.