Section 4
Polystat 4-3
Cole Parmer
To display/change the low temperature alarm settings, with the current
reservoir fluid temperature displayed, press
for two seconds. will
illuminate and the display will flash the current low limit value. If desired, use
the arrow keys to change the value. The low-end range is -13°C to 25°C.
NOTE
If the arrow keys are not pressed within 10 seconds the display will
return to the current reservoir temperature. You cannot set the alarm closer
than 2°C of the setpoint.
Press
again to confirm the new value. The display will rapidly flash the
new value for a short period and then will illuminate.
NOTE
If the value is not confirmed within 10 seconds the display will
return to the recirculating fluid temperature and ignore any changes.
With
illuminated the display will flash the current high limit value. If
desired, use the arrow keys to change the value. The high-end range is 10°C
to 35°C.
NOTE
If the arrow keys are not pressed within 10 seconds the display will
return to the current reservoir temperature. You cannot set the alarm closer
than 2°C of the setpoint.
Press
again to accept the new value.
The display will rapidly flash the
new value for a short period and return to the current reservoir temperature.
NOTE
If the value is not confirmed within 10 seconds the display will
return to the recirculating fluid temperature and ignore any changes.
Temp Alarms