Section 4
4-6 Polystat
Cole Parmer
To turn the chiller off press
.
When the display goes blank it is safe to place the circuit protector located on
the rear of the chiller to the down position.
NOTE
Starts per hour should be limited to no more than twelve.
Using any other means to shut the chiller down can reduce the life of
the compressor.
The circuit protector located on the rear of the chiller is not intended
to be used as a disconnecting means.
Always turn the chiller off and disconnect it from its supply voltage
before moving.
Stopping
CAUTION
CAUTION
CAUTION
Draining
Storage
Before using any fluid or performing maintenance where contact
with the fluid is likely refer to the manufacturer’s MSDS for handling
precautions.
The drain port, a 1/2" brass Female NPT plug, is located on the rear of the
chiller.
After draining we recommend inserting a wet-vac into the reservoir and
inlet/outlet lines to thoroughly remove any excess fluid.
Reinstall the plug.
CAUTION
CAUTION
If the chiller is to be transported and/or stored in cold temperatures
it needs to be drained and then flushed with a 50/50 laboratory grade
glycol/water mixture.
The chiller can be stored for up to 90 days inside the temperature range of
-25°C to +60°C (-13°F to +140°F).
If necessary, when removing the chiller from storage allow it time to warm
up and dry out in order to prevent any condensation issues.