Polystat 5-1
Cole Parmer
Cleaning
Condenser
Fluid Maintenance
Reservoir
Section 5
Preventive Maintenance
CAUTION
CAUTION
CAUTION
Laboratory Grade
Ethylene glycol (EG) is poisonous and flammable.
Before performing any preventive maintenance refer to the
manufacturer’s most current MSDS for handling precautions.
Disconnect the power cord prior to performing any maintenance.
Handle the chiller with care. Sudden jolts or drops can damage its
components.
There are no user serviceable components within the chiller cabinet.
Only Cole Parmer should provide any required replacement parts listed
below.
Clean the chiller's surface with a soft cloth and warm water only.
Failure to clean the condenser filter will cause a loss of cooling
capacity and lead to premature failure of the cooling system.
Clean the filter with a soft-bristle brush. When it is time for a more thorough
cleaning blow compressed air on the condenser to clean it.
The condenser fins are very sharp.
Check the fluid level, concentration and pH on a regular basis, see Section 3.
Change the fluid if it is discolored.
Periodically inspect the fluid inside the reservoir. If cleaning is necessary,
flush the reservoir with a cleaning fluid compatible with the circulating system
and the cooling fluid.
Failure to keep the reservoir full to the lip will result in
a loss of cooling capacity.