For the SBL-2D the following table indicates the heat input required to maintain a range of temperatures,
8.
with the energy regulator control:
Bath temperature
heater Switches
energy reg Setting
200°C (392°F)
LOW
7 to 8
300°C (572°F)
LOW
9 to 10
400°C (752°F)
MEDIUM
2 to 3
500°C (932°F)
MEDIUM
8 to 9
600°C (1112°F)
HIGH
4 to 5
The fourth heater is only used as a ‘boost’ heater.
THE FOURTH HEATER IS NOT CONTROLLED BY THE ENERGY REGULATOR AND SHOULD BE
SWITCHED OFF WHEN CONTROLL TEMPERATURE IS REACHED.
Maximum temperature for this bath: 600°C (1112°F).
Energy
Regulator
Heater
Selector
Switch
Heater
Selection
Switches
14