15
cLEANING AND DISINFEcTING
Thoroughly clean all components to remove medication residuals and possible impurities after each treatment.
The compressor (1) and the air hose (6) should be cleaned with a clean, moist cloth.
Always wash your hands well before cleaning and disinfecting the accessories.
Do not expose the compressor to water or heat.
Replace the air hose for each treatment with a new patient or in case of impurities.
Unplug the device before cleaning.
cleaning with water
•
Wash all nebuliser components
(except air hose)
under warm tap water (max. 60 °c) for about 5 mi-
nutes
adding if necessary a small quantity of detergent following dosage and use limitations as provided by
detergent manufacturer.
• Rinse thoroughly making sure that all deposits are washed away and leave to dry.
Disinfecting
• All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage
and use limitations as provided by disinfectant manufacturer.
• Disinfectants are usually available at pharmacies.
Sterilizing with steam
• All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.)
or 134 °C (7 min.).
• After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.