19
4. Instructions For Use Con’t.
4.7 Running a Cycle
To operate each cycle, follow these steps and watch the LCD .
RUBBER / PLASTIC (S)
121˚C / 15 min
RUBBER / PLASTIC (S)
121˚C / 30 min
Cycle you selected
CYCLE NUMBER 000000
As the cycle is running,
various sounds will be heard.
This is the normal functioning of the unit.
or
or
14:23 11/15/2006
SELECT A CYCLE
Cycle you selected
PRESS START
Cycle you selected
WARMING UP
CONDITIONING
110˚C
143 kPa
PRESSURIZING
125˚C
232 kPa
STERILIZING
136.8˚C 330 kPa 3:29
VENTING
108˚C
134 kPa
1. Turn the power switch
at the back of the unit to ON.
The LCD display reads:
2. Press the appropriate cycle button on the
keypad to scroll through the available cycles
The display will read either:
As the button is released the display reads:
The amber indicator light comes on. To indicate that the cycle is in progress, the following
messages are displayed as the cycle runs.
The unit’s cycle counter is displayed at the start, after the start button is depressed.
HOLLOW WRAPPED (S)
134˚C / 3.5 min
HOLLOW WRAPPED (S)
134˚C / 18 min
HOLLOW UNWRAPPED
(S) 134˚C / 3.5 min
HOLLOW UNWRAPPED
(S) 134˚C / 18 min
SOLID UNWRAPPED (N)
134˚C / 3.5 min