43
11. Specifications
11.1 Statim 2000S — Specifications
Machine Dimensions:
Length:
48.5 cm
Width:
41.5 cm
Height:
15 cm
Cassette Size (External):
Length:
41 cm (includes handles)
Width:
19.5 cm
Height:
4 cm
Cassette Size (Internal):
Length:
28 cm
Width:
18 cm
Height:
4 cm
Sterilization Chamber Volume:
1.8 L
Reservoir Volume:
4.0 L
Weight (Without water):
21 kg
Clearance required:
Top:
5 cm
Sides:
5 cm
Back:
5 cm
Front:
48 cm
Clearance required for movement of the door(s):
48 cm
Minimum charge in the water reservoir:
550 mL
PRV value (pressure relief valve):
Set at 43.5 PSI to release pressure
in overpressure situations
Thermal Fuse:
Cuts power to boiler in event of
overheat
Electrical Rating:
220 - 240 V, 50/60 Hz, 6 A
Current:
AC
Protection Class:
I
Protection: covered
Ambient Operating Temperature:
5˚C - 40˚C
Sound levels:
Mean - 56 dB, Peak - 65 dB
Humidity: 80% Max.
Max. Altitude:
2000 m
Max. Water Consumption:
268mL
Ambient Operating Pressure:
70kPa - 106kPa