ENGLISH
31
12
Electromagnetic
compatibility
Manufacturer’s guide and declaration – electromagnetic immunity
This camera is intended for use in the electromagnetic environment specified below.
The user must ensure that it is used in this environment.
Immunity test
IEC 60601
Level of gravity
Compliance
level
Electromagnetic environment - guide
Electrostatic
discharges
IEC 61000-4-2
± 6 kV in Contact
± 8 kV in air
± 6 kV
± 8 kV
The floor must be wooden, concrete or
tiled. If the floor is covered with a synthetic
material, the relative
humidity must be at least 30 %.
Rapid transient
peaks
IEC 61000-4-4
± 2 kV for power
lines
± 1 kV for input/
output lines
± 2 kV
± 1 kV
The quality of the main power supply
must be that of a commercial or typical
hospital environment.
Electric shocks
IEC 61000-4-5
Differential mode
± 1 kV
Shared mode
± 2 kV
± 1 kV
± 2 kV
The quality of the electric power supply
must be that of a commercial or typical
hospital environment.
Power dips,
power failures,
short power
interruptions
and variations in
power voltage
IEC 61000-4-11
• < 5 % UT –
for 10 ms
• 40 % UT –
for 100 ms
• 70 % UT –
for 500 ms
• < 5 % UT –
for 5 s
< 5 % UT
10 ms
< 40 % UT
100 ms
< 70 % UT
500 ms
< 5 % UT
5 s
The quality of the main power supply
must be that of a commercial or typical
hospital environment.
If the user of this camera requires it to
continue functioning during interruptions
of the main power supply, it is advisable
for this camera to be powered by an
inverter or a battery.
System
frequency
magnetic field
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The system frequency magnetic field
must be at a level that is characteristic of
location in a commercial or typical
hospital environment.
Summary of Contents for SYMBIOZ Cam+
Page 2: ...2...
Page 4: ...4...
Page 5: ...INTUITIVE CONNECTION English ENGLISH 5...
Page 6: ...ENGLISH 6 6...
Page 35: ...ENGLISH 35 13Symbols C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 36: ...36...
Page 37: ...INTUITIVE CONNECTION Fran ais FRAN AIS 37...
Page 38: ...FRAN AIS 38 38...
Page 67: ...67 FRAN AIS 13Symboles C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 68: ...68...
Page 69: ...INTUITIVE CONNECTION Espa ol ESPA OL 69...
Page 70: ...70 ESPA OL 70...
Page 99: ...99 ESPA OL 13S mbolos C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 100: ...100...
Page 101: ...INTUITIVE CONNECTION Italiano ITALIANO 101...
Page 102: ...102 ITALIANO 102...
Page 131: ...131 ITALIANO 13Simboli C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 132: ...132...
Page 133: ...INTUITIVE CONNECTION Deutsch DEUTSCH 133...
Page 134: ...134 DEUTSCH 134...
Page 163: ...163 DEUTSCH 13Symbole C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 164: ...164...
Page 165: ...INTUITIVE CONNECTION Nederlands NEDERLANDS 165...
Page 166: ...166 NEDERLANDS 166...
Page 195: ...195 NEDERLANDS 13Symbolen C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...
Page 196: ...196...
Page 197: ...INTUITIVE CONNECTION Polski POLSKI 197...
Page 198: ...198 POLSKI 198...
Page 227: ...227 POLSKI 13Symbole C1 C14 L1 S1 Z1 C2 C5 C10 C11 C12 C13 S2 C6 C3 C4 C7 C9 C8...