  IE-LOG-S3120E-03 

Instruction manual for use of temperature and RH logger S3120E 

Logger is designed for measurement and record of ambient temperature and relative humidity. 

Measuring sensors of temperature and humidity are attached to logger. Measured values including 
calculated  dew  point  temperature  are  displayed  on  the  two-line  LCD  display  and  are  stored  in 
selectable time interval to internal nonvolatile memory. All logger control and setting are performed 
from the PC and password is applicable. It is enabled to switch ON and OFF the logger by delivered 
magnet  (this  possibility can be in  configuration  disabled).  It  is  also  enabled to  program  automatic 
start  in  certain  day  and  time  (for  one  month  forward).  Start/stop  magnet  enables  also  to  clear  the 
minimum and maximum value memory. 

Minimum  and  maximum  measured  values  can  be  displayed  (display  switches  to  actual 

measured  values  and  min/max  values  automatically).  It  is  also  possible  to  operate  logger  with 
switched OFF display. Short display of actual measured values is enabled by means of magnet. 

Switched ON logger every 10 seconds (independently on logging interval) updates MIN/MAX 

memory,  compares  measured  values  of  each  quantity  with  two  adjustable  limits  for  each  quantity 
and  exceeding  of  limits  is  indicated  on  the  display  (alarm  function).  Also  memory  alarm  mode  is 
selectable, when alarm is indicated permanently till alarm memory reset. Alarm function is enabled 
or disabled for each quantity individually. 

Logging  mode  can  be  adjusted  as  non-cyclic,  when  logging  stops  after  filling  the  memory. 

In cyclic  mode  oldest  stored  values  are  overwritten  by  new.  In  addition  logging  mode  can  be 
selected when logging is active only if measured value is out of adjusted alarm limits. 

Stored  values  can  be  transferred  from  logger  memory  to  the  PC  by  means  of  communication 

adapter. Communication adapter can be connected to the logger permanently  – data logging is not 
interrupted even if data download appears.  

Logger evaluates minimum battery voltage and its drop below allowed limit is indicated on the 

display.  At  the  same  time  value  of  remaining  battery  capacity  is  available  by  means  of  the  PC 
program and appears on the logger LCD in % (every time after switch ON). 
 

Warning

 

The device can be services only by a qualified person. The device contains no serviceable parts 

inside. 

Don’t use the device if it doesn’t work correctly. If you think, that the device is not working 

correctly, let check it by qualified service person. 

It’s forbidden to use the device without the cover. Inside the device can be a dangerous voltage 

and may be risk of electric shock. 

 

Technical parameters: 

 

Measuring parameters: 
  Ambient temperature (RTD sensor Pt1000/3850ppm): 

 

Measuring range: -30 to +70 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ±0.6 °C from –30 to +30 °C, ±0.8 °C from +30 to +70 °C 

  Relative humidity (reading is temperature compensated at entire temperature range): 

 

Measuring range: 0 to 100 %RH 

   

 

Resolution: 0.1 %RH 

   

 

Accuracy: ± 3.0 %RH from 5 to 95 %RH at 23 °C 

  Dew point (value calculated from temperature and humidity): 

 

   

 

Range: -60 to +70 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ± 2.0 °C at ambient temperature T < 25°C and RV > 30 %, for 

 

 

 

     more details see Appendix A 

Response time with plastic sensor cover (air flow approximately 1 m/s): 
  temperature: t63 < 2 min, t90 < 8 min (temperature step 20 °C) 

Summary of Contents for S3120E

Page 1: ...nd improvement of all its products Manufacturer reserves the right to make technical changes without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the...

Page 2: ...ps after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm l...

Page 3: ...nterval 10 seconds min 1 year Notice the above lives are valid if logger is operated at temperature from 5 to 35 C If logger is operated often outside of the above temperature range life can be reduce...

Page 4: ...LCD reset memory of MIN MAX values if needed check free space in data memory optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger i...

Page 5: ...ties indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings are displayed up to estimated remaini...

Page 6: ...er replacement of totally discharged battery for new one State is indicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below...

Page 7: ...is necessary immediately till indication point is displayed to remove magnet from guide slots Reading CLR MIN MAX appears for several seconds and MIN MAX values will be cleared Battery replacement Low...

Page 8: ...elf after its life is necessary to liquidate ecologically End of operation Disconnect the device and dispose it according to current legislation for dealing with electronic equipment WEEE directive El...

Reviews: