IE-LOG-S3120E-03 

 

  3 

  relative humidity: t63 < 15 s, t90 < 50 s (humidity step 30 %RH, constant temperature)

 

Measuring interval, alarm evaluation and MIN/MAX memory update: 
  standard mode (no low-power mode): every 10 s 
  low-power mode: every 1 minute 
Logging interval to memory: 
  standard mode: 10 s to 24 h (20 steps) 
  low-power mode: 1 minute to 24 h (17 steps) 
Memory capacity:  

for non cyclic mode 

16 252  

for cyclic mode 

15 296 

  Specified values are maximum possible and can be reached only if record is not interrupted 

(since last memory erasing)  

Communication with computer: via RS232 (serial port) by means of COM adapter or USB 

port by means of USB adapter; data transfer from logger via communication adapter is 
optical 

Real time clock: adjustable from computer, integrated calendar including leap years 
Error of internal RTC: < 200 ppm (i.e. 0.02 %, 17.28 s in 24 h) 
Power: Lithium battery 3.6 V size AA 
Typical battery life: 
standard mode (data download to the PC approximately once a week): 2.5 years 
low-power mode (data download to the PC approximately once a week): 6 years 

            on-line mode with interval 1 minute: min. 1.5 years 
            on-line mode with interval 10 seconds: min. 1 year 

Notice: the above lives are valid if logger is operated at temperature from -5 to +35°C. If 

logger is operated often outside of the above temperature range life can be reduced to 75% 

Protection: IP30 
Operational conditions: 
  Operational temperature range: -30 to +70 °C 
  Operational humidity range: 0 to 100 %RH 
   Specification of outer characteristics accordingly to Czech National Standard 33-2000-3: 

normal environment accordingly to appendix NM:  AE1, AN1, AR1, BE1 

   Operational position: negligible  
  Logger installation: by self adhesive Dual Lock, applied to clean, flat surface 

Not allowed manipulation: it is not allowed to remove sensor cover and mechanically 

damage sensor under cover. Sensors of temperature and humidity should not get to 
direct contact with water or other fluids. 

Limit condition: temperature -40 to +70 °C, humidity 0 to 100 %RH 
Storing condition: temperature -40 to +85 °C, humidity 0 to 100 %RH 
Dimensions: 93 x 64 x 29 mm 
Weight including battery: approximately 115 g 
Material of the case: ABS 

 

Logger operation 

Logger 

comes 

complete 

with 

installed  battery  and  switched  OFF. 
Before  operation  it  is  necessary  by 
means  of  installed  user  PC  software  to 
set  logging  parameters  and  other 
features.  For  communication  with  the 
PC 

communication 

adapter 

is 

necessary (not included in delivery). For 

Summary of Contents for S3120E

Page 1: ...nd improvement of all its products Manufacturer reserves the right to make technical changes without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the...

Page 2: ...ps after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm l...

Page 3: ...nterval 10 seconds min 1 year Notice the above lives are valid if logger is operated at temperature from 5 to 35 C If logger is operated often outside of the above temperature range life can be reduce...

Page 4: ...LCD reset memory of MIN MAX values if needed check free space in data memory optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger i...

Page 5: ...ties indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings are displayed up to estimated remaini...

Page 6: ...er replacement of totally discharged battery for new one State is indicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below...

Page 7: ...is necessary immediately till indication point is displayed to remove magnet from guide slots Reading CLR MIN MAX appears for several seconds and MIN MAX values will be cleared Battery replacement Low...

Page 8: ...elf after its life is necessary to liquidate ecologically End of operation Disconnect the device and dispose it according to current legislation for dealing with electronic equipment WEEE directive El...

Reviews: