  IE-LOG-S3120E-03 

connection via RS232 serial port it is necessary to use COM  ADAPTER, for connection via USB 
port it is necessary to use USB ADAPTER. Connect adapter connector to proper computer port and 
plug the adapter to the guide slots on the side of logger.  

 

Notice:

 USB connector can be located also from the computer front side 

After connecting the logger to the computer reading of logger info is enabled by means of the 

PC software and also setting of the instrument accordingly to the user needs (menu 

Configuration / 

Setting of instrument  parameters

). Before logging start it is necessary: 

 

 

check or optionally set the logger real time clock 

 

select suitable logging interval 

 

select logging mode (cyclic or non cyclic) 

 

switch  ON  the  logger  (or  switch  OFF,  if  it  is  about  to  be  switched  ON  by  magnet  or 
automatically with delayed start) 

 

enable or disable the option to switch ON the logger by magnet 

 

enable or disable the option to switch OFF the logger by magnet 

 

enable or disable the option to clear minimum and maximum value memory by magnet 

 

set date and time of logger automatic switching ON logger or disable this option 

 

select if record runs permanently or only if alarm is active 

 

If  alarms  are  about  to  be  applied,  set  both  limits  for  each  measured  quantity  and  enable 
alarm 

 

optionally enable permanent alarm indication (alarm with memory) 

 

switch ON or switch OFF display logger 

 

optionally switch ON displaying of MIN/MAX values on the LCD 

 

reset memory of MIN/MAX values (if needed) 

 

check free space in data memory, optionally erase data memory of the logger 

 

enter password if protection against unauthorized manipulation with the logger is necessary  

 
Logging interval between to subsequent measurements is specified by the user. Memorizing of the 
first value is synchronized with the internal real time clock, so the logging is performed at sharp 
multiples of minutes, hours and days. E.g. after starting the logging with the 15 minute interval the 
first value is not stored immediately, but after the internal clock gets the status of a quarter, a half or 
a whole hour. After starting the logging with the 6 hour interval the first value is stored at that 
whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is 
performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging 
start. After communication with computer or after start by magnet logger automatically waits for 
the nearest whole multiple of time and then first measurement is performed. This is also necessary 
to take into account when setting time of automatic logger switch ON. 

Notice:

 If logger operates as permanently connected to the computer, using of magnet start/stop 

is disabled. 

 To  enable  the  logger  control  by  magnet  is  suitable  only  in  cases,  when  possibility  of 

unauthorized manipulation to the logger operation is eliminated. 

 

Reading on the display in usual operation (logger switched ON) 

 
After switching ON the logger all LCD symbols are displayed for checking 
the display. 
 
 
Then actual date and time in logger is displayed for approximately 4 s. 
 

Summary of Contents for S3120E

Page 1: ...nd improvement of all its products Manufacturer reserves the right to make technical changes without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the...

Page 2: ...ps after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm l...

Page 3: ...nterval 10 seconds min 1 year Notice the above lives are valid if logger is operated at temperature from 5 to 35 C If logger is operated often outside of the above temperature range life can be reduce...

Page 4: ...LCD reset memory of MIN MAX values if needed check free space in data memory optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger i...

Page 5: ...ties indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings are displayed up to estimated remaini...

Page 6: ...er replacement of totally discharged battery for new one State is indicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below...

Page 7: ...is necessary immediately till indication point is displayed to remove magnet from guide slots Reading CLR MIN MAX appears for several seconds and MIN MAX values will be cleared Battery replacement Low...

Page 8: ...elf after its life is necessary to liquidate ecologically End of operation Disconnect the device and dispose it according to current legislation for dealing with electronic equipment WEEE directive El...

Reviews: