IE-LOG-S3120E-03 

 

  5 

Consequently  reading  of  estimated  remaining  battery  capacity  for 
approximately  2  s  is  displayed  (values  0  to  100%).  It  is  valid  if  logger  is 
operated at temperature from -5 to +35°C. If logger is operated often outside 
of  the  above  temperature  range  battery  life  can  be  reduced  to  75%,  i.e.  if 
displayed remaining  battery capacity drops below 25%, it is  recommended 
to replace the battery. 

 

If display is switched ON

, actual reading of measured values is displayed -  

ambient temperature (°C) on the LCD upper line, relative humidity (%RH) 
on the LCD lower line. Symbol LOG indicates data logging in progress – if 
it blinks, data memory is filled up to more than 90%. 

 

Every 5 s display is automatically switched to displaying of other measured 
or  calculated  quantity.Logger  now  displays  ambient  temperature  and  dew 
point temperature (LCD line marked by symbol DP). 

 

Switched  ON  logger  permanently  (with  10  s  interval)  updates  memory  of 
minimum and maximum values of each measured (or calculated) quantity. If 
displaying of MIN/MAX values is selected, minimum measured values are 
displayed  step  by  step  (indicated  by  symbol  MIN)  and  then  similarly 
maximum  measured  values  of  all  quantities  (indicated  by  symbol  MAX). 
The  entire  cycle  is  repeated  periodically,  i.e.  reading  of  actual  measured 
values follows. 

 

If  display  is  switched  OFF,

  all  above  readings  are  displayed  up  to 

estimated remaining battery capacity and then display goes out. If logger is 
switched  ON  symbol  LOG  is  displayed  (it  blinks  if  memory  occupation  is 
higher than 90%). 

 

If display is OFF and logger is in mode when record runs only when alarm 
is  active, the LOG symbol  is  replaced by adjacent  symbol  "–"  (hyphen). It 
appears in case, all measured values are inside of adjusted alarm limits and 
data  logging  therefore  does  not  run.  Displayed  symbol  indicates  logger  is 
ON. 

 

If  information  on  actual  measured  values  is  needed,  it  is  possible  anytime  to  display  reading 

display  by  means  of  the  magnet  (only  if  communication  adapter  is not  connected  permanently). 
Plug magnet into guide slots from logger front side for approximately 4 s and wait till reading on 
the  display  appears.  If  logger  has  enabled  the  function  switch  OFF  by  magnet,  resp.  MIN/MAX 
memory clear by magnet, do not remove magnet from guide slots before decimal point symbol goes 
out – logger would be switched OFF, resp. MIN/MAX memory would be cleared! Display reading 
initialized by magnet goes out automatically after 30 s. Remove magnet from slots anytime during 
actual reading is ON or later. 

 
 

Temporary displaying of actual reading by magnet 

 

 

Summary of Contents for S3120E

Page 1: ...nd improvement of all its products Manufacturer reserves the right to make technical changes without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the...

Page 2: ...ps after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm l...

Page 3: ...nterval 10 seconds min 1 year Notice the above lives are valid if logger is operated at temperature from 5 to 35 C If logger is operated often outside of the above temperature range life can be reduce...

Page 4: ...LCD reset memory of MIN MAX values if needed check free space in data memory optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger i...

Page 5: ...ties indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings are displayed up to estimated remaini...

Page 6: ...er replacement of totally discharged battery for new one State is indicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below...

Page 7: ...is necessary immediately till indication point is displayed to remove magnet from guide slots Reading CLR MIN MAX appears for several seconds and MIN MAX values will be cleared Battery replacement Low...

Page 8: ...elf after its life is necessary to liquidate ecologically End of operation Disconnect the device and dispose it according to current legislation for dealing with electronic equipment WEEE directive El...

Reviews: