IE-LOG-S3120E-03 

 

  7 

 

 

 

 

Switching the logger OFF by magnet 

The  procedure  is  identical  with  the  above  procedure  for  switching  ON.  If  decimal  point  does 

not appear after 1 s, it is necessary remove the magnet and repeat the procedure. 

 

 

Reset of MIN/MAX values by magnet 

Function enables to clear MIN/MAX values by magnet without using computer. It is necessary 

to enable the function from the PC software before. 

Notice: 

It  is  not  possible  to  combine  this  function  with  function  of  switching  logger  OFF  by 

magnet! User software enables to select only one of them (or none). 

Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point 

appears right  on  LCD upper line.  After decimal  point appearance it is  necessary immediately (till 
indication point is displayed) to remove magnet from guide slots. Reading CLR MIN MAX appears 
for several seconds and MIN/MAX values will be cleared. 
 

 

 
 

Battery replacement 

Low  battery  is  indicated  on  the  display  by  blinking  of  reading  "BAT".  It  can  be  displayed 

permanently,  if  battery  voltage  is  too  low.  Replace  the  battery  for  new  one.  If  logger  is  operated 
often  in  temperature  below  -5°C  or  over  +35°C  and  PC  program  indicates  remaining  battery 
capacity below 25% it is also recommended to replace the battery. Applied is Lithium battery 3.6 V, 
size AA. Battery is located under logger lid. 

Warning:

  near  battery  fragile  glass  reed  contact  is  located  –  be  careful  not  to  damage  it.  Be 

careful in battery replacement! 

Replacement procedure: 

 

switch off the logger by the PC program or by magnet (if low battery allows) 

 

unscrew four corner screws and remove the lid 

 

remove old battery by pulling the glued tape 

Summary of Contents for S3120E

Page 1: ...nd improvement of all its products Manufacturer reserves the right to make technical changes without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the...

Page 2: ...ps after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm l...

Page 3: ...nterval 10 seconds min 1 year Notice the above lives are valid if logger is operated at temperature from 5 to 35 C If logger is operated often outside of the above temperature range life can be reduce...

Page 4: ...LCD reset memory of MIN MAX values if needed check free space in data memory optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger i...

Page 5: ...ties indicated by symbol MAX The entire cycle is repeated periodically i e reading of actual measured values follows If display is switched OFF all above readings are displayed up to estimated remaini...

Page 6: ...er replacement of totally discharged battery for new one State is indicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below...

Page 7: ...is necessary immediately till indication point is displayed to remove magnet from guide slots Reading CLR MIN MAX appears for several seconds and MIN MAX values will be cleared Battery replacement Low...

Page 8: ...elf after its life is necessary to liquidate ecologically End of operation Disconnect the device and dispose it according to current legislation for dealing with electronic equipment WEEE directive El...

Reviews: