background image

 
 

 

 

 

12

 

 

IE-COM-C4141-13

 

 

 

Reading  ALHI  indicates  adjusted  upper  limit  of  the  same  alarm 
(here air temperature). Set the desired value in the same way as in 
above  lower  limit.  If  needed  it  is  possible  to  get  back  to  lower  limit 
setting of the same alarm by pressing DISPLAY key. Press ENTER 
key to confirm new upper limit. 
  
 
Then you are offered to set alarm of other input value (here relative 
humidity). The procedure is the same as the above temperature limit 
setting.  Alarm  activation  of  each  value  can  be  disabled  by  setting 
lower  alarm  limit  of  the  desired  value  up  to  its  maximum.  This  is 
indicated by OFF reading at the position of numeric value. Value of 
upper limit of the same alarm is indifferent. 
 
It  is  possible  to  leave  the  alarm  setting  mode  by  pressing  MENU 
key.  New  adjusted  limits  up  to  pressing  MENU  key  are  stored 
in memory. 

 

 
 
 
 
 
 
 

 

 

Summary of Contents for Commeter C4141

Page 1: ...www cometsystem com Instruction manual Commeter C4141 Digital thermo hygro barometer with external probe...

Page 2: ...PERATION OF THE DEVICE 8 Switching ON and OFF the device 8 Displaying of actual measured values 8 Function HOLD storing of actual measured values and minimum a maximum memory 9 Functions and settings...

Page 3: ...emperature is measured by RTD sensor Ni1000 6180ppm Instrument is equipped with an external temperature relative humidity probe Instrument compares all measured values except pressure tendency with tw...

Page 4: ...tep 20 C Relative humidity Measuring range 0 to 100 RH Resolution 0 1 RH Accuracy 2 5 RH from 5 to 95 RH at 23 C Response time t90 1 min with plastic sensor cover without filter t90 1 min with plastic...

Page 5: ...5 s in case measured values remain stable Refresh interval is shortened to approximately 0 7 s only if measured values change Current consumption in this mode in usual operation is lower battery life...

Page 6: ...on device 0 to 100 RH probe Temperature and rel humidity restriction The measuring range of the external probe is limited accordance with the graph and must not be exceeded Storage temperature range 1...

Page 7: ...IE COM C4141 13 7 Dimensions...

Page 8: ...t the probe to avoid storing of improper values to the min max memory Displaying of actual measured values In this mode is instrument any time after switching ON It is possible to enter this mode from...

Page 9: ...ermanently updates minimum and maximum memory of each measured values Press MIN key resp MAX key in the default mode to display minimum resp maximum reading These minimum and maximum readings are indi...

Page 10: ...This value illustrates battery condition This selection by pressing ENTER key enables to enter mode of pressure correction to the sea level see below Display refresh mode is indicated In the FAST mode...

Page 11: ...nd setting It is possible to set lower and upper limit for each measured quantity Breaking of the limit is indicated by blinking of the appropriate value on the display If a new alarm was indicated i...

Page 12: ...hen you are offered to set alarm of other input value here relative humidity The procedure is the same as the above temperature limit setting Alarm activation of each value can be disabled by setting...

Page 13: ...replace battery with instrument switched OFF otherwise setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason do not disconnect the batter...

Page 14: ...14 IE COM C4141 13 Technical support and service Technical support and service is provided by distributor Contact is included in warranty certificate...