IE-COM-C4141-13
13
Battery replacement
Low battery voltage is indicated on the display with blinking reading "BAT". It is necessary to
replace it with new one as soon as possible. Battery is located under small cover on the
instrument lower side. It is absolutely necessary to replace battery with instrument switched
OFF, otherwise setting of d.REF. and AUDI (from menu selections) and data in memory
HOLD, MIN and MAX will be lost. For the same reason do not disconnect the battery for longer
than 1 minute even if instrument is switched OFF. If it happens (or if battery is totally
discharged), it is necessary to set again in appropriate menu selection LCD refreshment mode
(d.REF.), alarm audio indication (AUDI) and clear the minimum and maximum memory (CLR).