Logging mode 

 

 

In logging mode measured values are stored to instrument internal memory. Memory is non volatile, 
i.e. if battery voltage drops or battery is disconnected, stored data will not be lost. 
 
Device can work in two logging modes: 

 

Automatic mode

 

– actual measured values are stored to memory in adjustable time interval. 

Storing of the first value is synchronized with internal real time clock, so the memorizing is 
performed  at  sharp  multiples  of  minutes,  hours  and  days.  E.g.  if  the  storing  interval  is 
15 minutes value will be stored each 15., 30., 45. a 00. minute and zero second in each hour, 
if interval is 3 hours values will be stored in 00:00:00, 03:00:00, 06:00:00. etc. It means after 
starting automatic logging instrument waits for the nearest integer multiple of time and then 
stores the first value.  

Manual mode

 

– actual measured value is stored after pressing ENTER key (instrument should 

be in default mode of displaying actual measured values). Each storing is indicated by a 
short beep and position of the record in memory is displayed (reading MEMO with the record 
serial number). In this mode it is possible to store max. 1000 records.  

 

It is possible to select one of the three logging modes (MODE): 
 

Automatic non cyclic logging

, when logging stops after fulfilling the memory. Reading MEMO 

FULL is displayed (reading disappears after pressing any key) 

 

Automatic cyclic logging

, when after fulfilling the memory oldest data are replaced with new 

values  

 

Manual logging

, designed for manual storing of measured values by ENTER key  

  
Logging  modes  are  not  mutually  compatible  and  after  changing  logging  mode  memory  of  the 
instrument should be cleared. 

 

 

Logging in progress will always be interrupted by instrument connection to the computer or switching 
OFF  the  instrument  by  ON/OFF  key.  Logging  in  progress  will  be  automatically  stopped  if  battery 
voltage below 7 V is indicated by the instrument. This prevents to log values effected by low battery 
voltage.  
 

 
 

 

 

Logging start from device keyboard 

 
 

Start logging by confirmation of the menu item START (symbol Strt) 
by pressing ENTER key. 
 
Notice:  it  is  not  possible  to  start  logging  if  low  battery  voltage  is 
indicated. 

 

 

 

 

Summary of Contents for Commeter D0221

Page 1: ...www cometsystem com Instruction manual Commeter D0221 Dual channel thermometer and datalogger for RTD Pt1000 and Ni1000 6180ppm probes...

Page 2: ...alues and minimum a maximum memory 9 Functions and settings available from menu 9 Alarm indication and setting 11 LOGGING MODE 13 Logging start from device keyboard 13 Logging start with clearing memo...

Page 3: ...nd it is possible to store values into the non volatile memory Then it is possible to transfer data to a personal computer Instrument compares measured values of temperature with two adjustable levels...

Page 4: ...Pt1000 3850 The device with Ni1000 6180 probes Measuring range 50 to 250 C with the limitation according to the used probe Resolution 0 1 C Accuracy 0 4 C from 50 to 100 C 0 5 from reading from 100 t...

Page 5: ...ation without ac dc adapter plugged Accumulator works only as a standby source in case of power mains failure Average current consumption 0 08 to 0 3 mA depending on operation mode ___________________...

Page 6: ...Backlighting turns off automatically 10 seconds after the last keystroke Protection IP20 Dimensions 146 x 71 x 27 mm Weight Approximately 150 g including battery without probe Operation conditions Am...

Page 7: ...IE COM D0221 05 7 Dimensions...

Page 8: ...ent storing of incorrect value to minimum and maximum memory If a voltage drop below critical limit occurred since last adjustment of internal real time clock or battery disconnection longer than appr...

Page 9: ...value from MENU see below Each pressing of the HOLD key in the default mode causes value stored in HOLD memory is replaced with actual temperature Switched ON instrument permanently updates minimum an...

Page 10: ...witched on On or switched off OFF Press ENTER key to change actual setting Notice if the battery voltage is low audio indication is out of operation to reduce current consumption independently on this...

Page 11: ...ch pressing of ENTER key causes displaying information on storing interval actual logging mode MODE memory occupation MEMO in automatic logging mode memory occupation in is displayed in manual mode th...

Page 12: ...pressing DISPLAY key Press ENTER key to confirm new upper limit Then you are offered to set alarm of other input value here temperature of channel 2 The procedure is the same as the above temperature...

Page 13: ...dicated by a short beep and position of the record in memory is displayed reading MEMO with the record serial number In this mode it is possible to store max 1000 records It is possible to select one...

Page 14: ...indicated with reading LOG on display If memory is filled over 90 symbol LOG is blinking only in automatic logging Logging start with clearing memory and logging mode selection It is possible to clear...

Page 15: ...de is selected after confirmation by ENTER key logging is active In automatic logging mode i e cyclic or non cyclic last applied storing interval is offered to confirm or modify See above procedure Lo...

Page 16: ...protocol of measurement including instrument serial number and a table of logged values with date and time After disconnection of the communication cable reading PC disappears and instrument is in the...

Page 17: ...t for automatic Installator run If Installator is not run automatically depends on setting of operating system find in file browser file CDSetup exe in main directory of the installation CD and click...

Page 18: ...le of measured values occurs Detailed information on file is available from context menu menu occurs after clicking right mouse button on the position of table In context menu function export to text...

Page 19: ...cessary to replace battery with instrument switched OFF otherwise internal time and setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason...

Page 20: ...20 IE COM D0221 05 Technical support and service Technical support and service is provided by distributor Contact is included in warranty certificate...

Reviews: