48
IE-MLG-M1200E-01
Appendixes
Appendix 1: Selected error messages
Error
Description and solution
Error 1
The value is out of the input range (low limitation).
- shorted Pt1000/M probe
- negative input voltage (thermocouples, 0 to 10 V)
- reverse input current (for 0 mA to 20 mA range)
Check the connected probes and signals!
Error 2
The value is out of the input range (high limitation).
- disconnected Pt1000/M probe
- too high input voltage (thermocouples, 0 to 10 V)
- too high input current (for 0 mA to 20 mA range)
Check the connected probes and signals!
Error 4
Digi_/M probe communication error.
Check the connected probe - it is probably not connected or it
is not Digi_/M line probe!
Error 5
Digi_/M probe communication error.
The probe is probably damaged. Try another probe!
Error 7
Pt1000/M probe communication error.
Check the connected probe - it is probably not connected or it
is not Pt1000/M line probe!
Error 8
Cold junction temperature measurement error.
The Multilogger is outside of the range of operational
temperatures or the device is damaged.
Error 9
The value is out of the measuring range (for example,
temperature from RTD and from the thermocouple probes).
The probe could be defective, the thermocouple type could
be different and there may be an undesirable coupling among
inputs or incorrect calibration of the Multilogger.
Try a
different probe, check the cables and unwanted coupling.
Error 10
The pressure sensor is not available.
Send the Multilogger for repair.
Error 11
The counter does not have valid data, the Multilogger has
been without batteries.
Erase the counter using the SW and repeat measurement.